Key result
Oscillometric noninvasive monitoring systematically underestimates invasive MAP by ~9 mmHg in anesthetized pumas.
Why the study?
Agreement between noninvasive oscillometric blood pressure monitoring and invasive blood pressure measurements in anesthetized pumas was not established.
Does noninvasive blood pressure obtained by an automated oscillometric monitor accurately agree with invasive blood pressure measurements in anesthetized pumas?
Does noninvasive blood pressure obtained by an automated oscillometric monitor accurately agree with invasive blood pressure measurements in anesthetized pumas?
Mean Difference: -8.7 (95% CI -32.1–14.7)
Automated oscillometric noninvasive blood pressure monitoring systematically underestimates invasive blood pressure and cannot be used interchangeably in anesthetized pumas.
OBJECTIVE: To evaluate agreement between noninvasive blood pressure (NIBP) obtained using an automated oscillometric blood pressure monitor and invasive blood pressure (IBP) measurements in anesthetized pumas (Puma concolor). STUDY DESIGN: Prospective experimental trial. ANIMALS: A group of five healthy adult captive pumas. METHODS: Pumas were anesthetized by intramuscular injection of ketamine, midazolam, dexmedetomidine, and methadone, followed by total intravenous anesthesia with propofol, remifentanil, ketamine, and lidocaine. NIBP was measured at the tail base using an oscillometric device, whereas IBP was obtained via dorsal pedal artery catheterization. Paired systolic (SAP), mean (MAP), and diastolic (DAP)_arterial blood pressures measurements were recorded every 5 minutes. Agreement was assessed using repeated-measures Bland-Altman analysis accounting for within-animal clustering. RESULTS: A total of 163 paired measurements were analyzed. Noninvasive measurements systematically underestimated IBP for SAP, MAP, and DAP. Bias (95% limits of agreement) was -11.8 (-41.1 to +17.4) mmHg for SAP, -8.7 (-32.1 to +14.7) mmHg for MAP, and -8.8 (-35.1 to +17.5) mmHg for DAP. Proportional bias was identified for SAP but not for MAP or DAP. The percentages of measurements that differed by ≤10 mmHg were 57.1% for MAP, 53.9% for DAP, and 41.1% for SAP. Measurements differing by ≤20 mmHg were 88.2% for MAP, 82.8% for DAP, and 73.6% for SAP. CONCLUSION AND CLINICAL RELEVANCE: The evaluated oscillometric device cannot be considered interchangeable with IBP monitoring in anesthetized pumas. Of the arterial blood pressure variables, MAP showed the most consistent agreement with IBP, although individual measurements may be associated with clinically relevant error.
No takes yet. Share an insight, caveat, or question.
Ribeiro et al. (2026) studied Anesthetized pumas (n=5). Automated oscillometric blood pressure monitor (NIBP) vs. Invasive blood pressure (IBP) measurements was evaluated on Agreement between noninvasive and invasive blood pressure measurements (Bias (mean difference) for MAP -8.7 mmHg, 95% CI -32.1 to 14.7). Noninvasive blood pressure monitoring using an automated oscillometric device systematically underestimated invasive blood pressure in anesthetized pumas (bias for MAP -8.7 mmHg).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: