Key result
7T cardiac MRI provides comparable strain and only ~10% lower LV volumes versus 3T.
Why the study?
Cardiac MRI at 7 Tesla is not clinically established due to spatial heterogeneities of transmit B1 magnitude and prolonged examination times.
Does cardiac MRI at 7T using a single B1 shim set provide comparable volumetric and strain parameters to 3T MRI in healthy volunteers?
Observational (n=10)
No
Does cardiac MRI at 7T using a single B1 shim set provide comparable volumetric and strain parameters to 3T MRI in healthy volunteers?
Cardiac MRI at 7T with a static B1 shimming workflow is feasible and provides comparable strain analysis and clinically acceptable volumetric quantification compared to standard 3T MRI.
Due to spatial heterogeneities of transmit B1 magnitude, cardiac MRI at 7 Tesla is still far from being clinically established. Different strategies were proposed to mitigate signal variations that may lead to a significant increase of the examination time. The aim of this study was to establish an efficient workflow for B1 shimming for cine imaging at 7T with a single B1 shim set and the investigation of the impact of field strength on volumetric and strain parameters Long- and short-axis images were acquired in ten volunteers at 3T and 7T. At 7T, B1 maps were acquired to calculate a single B1 phases him vector that was applied for the entire exam. Image quality reading was performed by two radiologists. Biventricular volumetric parameters were assessed by two readers. Biventricular function was further assessed for peak strain using feature tracking Image quality and contrast of long-axis images acquired at 7T was inferior compared to 3T, although all images were deemed to bediagnostically usable. In comparison to 3T, 7T slightly underestimated left ventricular volumes, with differences remaining under 10%of absolute values and thus of minimal clinical relevance. Strain parameters were not impacted by field strength A single B1 shim set provides excitation patterns without substantial signal dropouts in the heart. Moderate residual B1 variations and the reduced myocardium–blood contrast at 7T hindered neither quantification nor diagnostic quality. In conclusion, an efficient workflow could be realized at 7T for cardiac cine imaging for function and strain analysis SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at https://doi.org/10.1007/s10554-026-03741-6.
No takes yet. Share an insight, caveat, or question.
Jung et al. (2026) conducted an observational in Healthy volunteers (n=10). 7T cardiac MRI with static B1 shimming vs. 3T cardiac MRI was evaluated on Biventricular volumetric parameters, strain parameters, and image quality. 7T cardiac MRI with a single B1 shim set yielded diagnostically usable images with comparable strain parameters and only slightly underestimated left ventricular volumes (<10%) compared to 3T MRI.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: