Why the study?
Epicardial adipose tissue may limit effective radiofrequency ablation, and challenges exist in delivering effective lesions and managing device sensing in EV-ICD recipients undergoing ventricular tachycardia ablation.
Does hybrid endo-epicardial pulsed-field ablation combined with lead repositioning improve arrhythmia control and device function in an EV-ICD recipient with recurrent VT?
Case Report (n=1)
Does hybrid endo-epicardial pulsed-field ablation combined with lead repositioning improve arrhythmia control and device function in an EV-ICD recipient with recurrent VT?
A hybrid endo-epicardial pulsed-field ablation approach paired with lead repositioning is a feasible strategy to manage recurrent VT and device sensing issues in patients with extravascular ICDs.
BACKGROUND: Epicardial adipose (EA) tissue may limit effective radiofrequency ablation (RFA). OBJECTIVES: We sought to evaluate the lesion formation of different ablation strategies on ventricular myocardium with overlying EA. METHODS: Bovine myocardium with EA was placed in a circulating saline bath in an ex vivo model. Open-irrigated (OI) RFA was performed, parallel to the myocardium, over fat at 50 W for variable RF durations, variable contact force, catheter configurations (unipolar RF vs bipolar RF), and catheter irrigants (normal saline vs half-normal saline). Ablation was also performed with a needle-tipped ablation catheter (NTAC), perpendicular to the myocardium. RESULTS: Increasingly thick EA attenuated lesion size regardless of ablation strategy. RF applied with longer durations and increasing CF produced larger lesion volumes and deeper lesions with ablation over EA more than 3 mm but was unable to produce measurable lesions when EA less than 3 mm. Similarly, ablation with half normal saline irrigant created slightly deeper lesions than bipolar RF and unipolar RF with normal saline as EA thickness increased, but was unable to produce measurable lesions when EA more than 3 mm. Of all ablation strategies, only NTAC produced effective lesion volumes when ablating over thick (>3 mm) EA. CONCLUSIONS: While EA attenuates lesion depth and size, relatively larger, and deeper lesions can be achieved with longer RFA duration, higher CF, half normal saline irrigant, and, to a greater extent, by utilizing bipolar RF or NTAC, but only over thin adipose (<3 mm). Of those catheters/strategies tested, only NTAC was able to effectively deliver RF over thick (>3 mm) EA with this model.
No takes yet. Share an insight, caveat, or question.
Zipse et al. (2020) conducted a case report in Recurrent sustained monomorphic ventricular tachycardia in an EV-ICD recipient (n=1). Hybrid endo-epicardial pulsed-field ventricular tachycardia ablation and EV-ICD lead repositioning was evaluated on Acute arrhythmia termination and 30-day event-free survival.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: