Answers are grounded in this paper and cite their sources.
A rapid and highly sensitive ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) assay was developed for quantification of 7 antibiotics in low sample volumes (50 μL): amoxicillin, azithromycin, cefotaxime, ciprofloxacin, meropenem, metronidazole and piperacillin, for both routine monitoring and pharmacokinetic studies. After protein precipitation by acetonitrile, the antibiotics were separated on an Acquity UPLC HSS T3 column (run time, 4 min). The mobile phase consisted of a mixture of (A) ammonium acetate (pH 2.4; 5 mM) and (B) acetonitrile acidified with 0.1% formic acid, delivered at 500 μl/min in a gradient elution mode. Total time run was 2.75 min. Ions were detected in the turbo-ion-spray-positive and multiple-reaction-monitoring modes. The assay was accurate and reproductible for the quantification of the seven antibiotics in serum samples over large concentration ranges.
No takes yet. Share an insight, caveat, or question.
Magréault et al. (2019) studied this question.