Key result
AI model using ECG and phonocardiography detects moderate-or-greater aortic stenosis with ~93% sensitivity.
Why the study?
Valvular heart disease, particularly aortic stenosis, often remains asymptomatic until advanced stages, and traditional cardiac auscultation has limited diagnostic accuracy for early detection.
Does a deep learning model using phonocardiogram and single-lead ECG accurately detect moderate-or-greater aortic stenosis in patients undergoing routine echocardiography?
Cross-Sectional (n=951)
Blinded reference standard
Yes
Does a deep learning model using phonocardiogram and single-lead ECG accurately detect moderate-or-greater aortic stenosis in patients undergoing routine echocardiography?
Effect estimate: Sensitivity 92.9% (95% CI 86.4-98.4)
An AI-based model combining short-duration phonocardiograms and single-lead ECGs demonstrated high diagnostic performance for detecting clinically significant aortic stenosis, potentially serving as an effective screening tool to guide echocardiography referrals.
May aid noninvasive AS screening; extends AI phonocardiography validation but leaves open clinical adoption pending trials.
Background Valvular heart disease, particularly aortic stenosis (AS), often remains asymptomatic until advanced stages, making early detection challenging. Although cardiac auscultation is a traditional screening tool, its diagnostic accuracy is limited. We aimed to externally validate a deep learning–based model (eAS model) using phonocardiogram (PCG) and single‐lead electrocardiogram (ECG) signals to detect moderate‐or‐greater AS. Methods PCG and ECG recordings were acquired using a PCG device (Cardio‐EGG). Transthoracic echocardiography served as the reference standard. The external validation cohort consisted of 951 patients from 3 medical institutions, including 70 with moderate‐or‐greater AS. The primary outcome was sensitivity for detecting moderate‐or‐greater AS; secondary outcomes included the area under the receiver operating characteristic curve (AUROC), specificity, predictive values, and likelihood ratios (LR). Results The external validation cohort ( n = 951, median age 73 years, male 65%) exhibited more advanced valvular disease than the training/development cohort. The eAS model achieved an AUROC of 0.9632 (95% CI: 0.9471–0.9774). Using the pre‐determined threshold, sensitivity was 92.9% and specificity 87.9%, yielding a positive LR of 7.65 and negative LR of 0.08. Model‐derived probabilities showed strong associations with echocardiographic measures of AS severity (adjusted R 2 = 0.5051 for peak transvalvular velocity). Exploratory subgroup analyses showed generally consistent performance, although estimates were imprecise in small subgroups. Conclusion The eAS model demonstrated high diagnostic performance for detecting moderate‐or‐greater AS in a multicenter external validation cohort. AI‐based PCG/ECG analysis may support the decision to refer patients from auscultation to echocardiography for AS evaluation.
No takes yet. Share an insight, caveat, or question.
Saito et al. (2026) conducted a cross-sectional in Moderate-or-greater Aortic Stenosis (AS) (n=951). eAS deep learning model (PCG and single-lead ECG signals) vs. Transthoracic echocardiography (reference standard) was evaluated on Sensitivity for detecting moderate-or-greater aortic stenosis (Sensitivity 92.9%, 95% CI 86.4-98.4). The eAS deep learning model using phonocardiogram and single-lead ECG signals detected moderate-or-greater aortic stenosis with 92.9% sensitivity and an AUROC of 0.9632.