Efficient removal of middle-to-large molecules, such as α 1 -microglobulin (α 1 -MG, 33 kDa), is essential for mitigating long-term complications in patients undergoing hemodiafiltration (HDF). However, objective evaluation of diafilter performance requires standardized laboratory systems that can reproduce clinical transport characteristics while monitoring albumin leakage. This study established an in vitro experimental system utilizing recombinant α 1 -MG to evaluate post-dilution on-line HDF. Two configurations were compared: Circuit#1, featuring an auxiliary pump at the blood outlet, and Circuit#2, where the pump was removed to optimize flow regulation. Results demonstrated that Circuit#1 significantly deviated from clinical benchmarks due to internal pressure accumulation. Conversely, Circuit#2 successfully mirrored clinical trends, showing a strong positive correlation between α 1 -MG removal and albumin leakage. Observed clearances in Circuit#2 were slightly higher than clinical averages, likely due to the smaller molecular weight of the recombinant solute (31 vs. 33 kDa), reduced circulation volume, and the absence of α 1 in vitro -MG-IgA dimer binding complexes. This validated model provides a robust and reliable platform for the standardized assessment and development of next-generation blood purification devices.