Additively manufactured implants may be vulnerable to localized corrosion within narrow gaps, yet the critical crevice geometry and the effectiveness of multilayer coatings under such conditions remain insufficiently understood. This study compares the crevice-corrosion behavior of direct metal laser sintering (DMLS) and conventionally wrought mill-annealed Ti-6Al-4V alloys and evaluates a hierarchical coating comprising a plasma electrolytic oxidation (PEO) layer, a polycaprolactone (PCL) seal, and a porous polylactic acid (PLA) breath-figure topcoat. Electrochemical testing was conducted in simulated body fluid (SBF, pH 7.4, 37 °C) using a tapered crevice cell, followed by microscopic, compositional, and topographical analyses. Both uncoated alloys exhibited passive film breakdown; however, the wrought alloy subsequently repassivated, whereas the DMLS alloy underwent sustained localized attack within the critical crevice gap of ~11.4 µm. The transpassive potentials for DMLS and wrought Ti-6Al-4V with PEO/PCL/PLA coating system were 1.32 V and 1.45 V, respectively. Despite the intrinsic porosity and interfacial gaps of the coating, no crevice-type damage was detected in DMLS Ti-6Al-4V. PCL/PLA hierarchy increased the PEO pore resistance to ~3.5 kΩ cm2, reduced the passive current density by ~2 times, and increased the PEO barrier layer resistance to 107 Ω cm2.