Key result
Noninvasive intracardiac telemetry accurately reproduces electrograms compared with standard multichannel physiological recorders.
Why the study?
The feasibility of noninvasive intracardiac telemetry for dysrhythmia analysis had not been demonstrated.
Answers are grounded in this paper and cite their sources.
Noninvasive intracardiac telemetry is feasible in EP studies; leaves open routine adoption pending prospective validation.
In order to demonstrate the feasibility of dysrhythmia analysis by means of noninvasive intracardiac telemetry, a custom‐modified AFP DDD pulse generator was attached to temporary pacing electrodes in 15 patients undergoing electrophysiology studies for a variety oftacliy‐ or bradyarrhythmias. Intracardiac atrial and ventricular electrograms were obtained during induction, termination, and drug intervention in patients with these arrhythmias. The telemetered computer‐generated hardcopy tracings compared favorably to those obtained through a standard multichannel physiological recorder, and thus confirmed the accuracy of the technique. Therefore, microprocessor‐based telemetry ofintracardiac electrograms may have a prominent role in the design of multifunction devices for both bradycardias and tachycardias
No takes yet. Share an insight, caveat, or question.
Luceri et al. (1987) studied Tachyarrhythmias or bradyarrhythmias (n=15). Custom-modified AFP DDD pulse generator for noninvasive intracardiac telemetry vs. Standard multichannel physiological recorder was evaluated on Accuracy of telemetered computer-generated hardcopy tracings compared to standard recording. Noninvasive intracardiac telemetry using a custom-modified pulse generator produced accurate electrograms that compared favorably to a standard multichannel physiological recorder.
Synapse has enriched 10 related papers. Consider them for comparative context: