Key result
Benchmarking wearable arrhythmia detectors requires 54 clusters to detect a ~5-point change in false alarms.
Why the study?
Wearable-triggered arrhythmia detectors are benchmarked against reference annotations with uncertain atrial fibrillation episode boundaries, and the effect of this boundary ambiguity on reported false-alarm performance has not been characterised.
Provides a boundary-aware benchmarking protocol and statistical design parameters to separate wearable arrhythmia detector error from reference annotation error when evaluating false-alarm performance.
Boundary ambiguity may distort reported false-alarm rates in wearable AF detectors; simulation framework leaves open the magnitude of bias in practice.
Wearable-triggered arrhythmia detectors are benchmarked against reference annotations whose atrial fibrillation episode boundaries are themselves uncertain, yet the effect of that boundary ambiguity on reported false-alarm performance has not been characterised. The framework is then embedded in a before-after-control-impact design with a declared variance structure (between-cluster SD = 6.0, residual SD = 4.5, intraclass correlation = 0.64) and evaluated across 2000 replications per design cell for cluster counts of 8, 16, 24 and 32 and effect sizes of 3, 5 and 8 percentage points.54 clusters are required to detect a 5 percentage point change at 80% power, and that the minimum detectable effect at 80% power falls steeply as clusters increase from 8 to 32. The contribution is a boundary-aware benchmarking protocol that separates detector error from annotation error, together with the design parameters needed to evaluate it credibly.
No takes yet. Share an insight, caveat, or question.
Hoareau et al. (2026) studied Atrial fibrillation. Boundary-aware benchmarking protocol for wearable-triggered arrhythmia detectors vs. Standard reference annotations with episode boundary ambiguity was evaluated on False-alarm performance under episode boundary ambiguity (minimum detectable effect at 80% power). Boundary-aware benchmarking of wearable arrhythmia detectors requires 54 clusters to detect a 5 percentage point false-alarm change at 80% power (ICC=0.64).