Key result
Proximal LBBP with a 4.7-F lumenless ICD lead causes ~6% partial atrial coil extension.
Why the study?
Complete accommodation of the integrated bipolar coil of a 4.7-F lumenless defibrillator lead within the RV can be challenging during LBBP, particularly with proximal LBBP where shorter lead-to-HB distance is associated with greater LVEF improvement.
Describes a technique for slack formation during proximal left bundle branch pacing to facilitate complete accommodation of a 4.7-F lumenless defibrillation lead's integrated bipolar coil within the right ventricle.
A 4.7-F lumenless lead (OmniaSecure, Medtronic) enables left bundle branch pacing (LBBP) with a defibrillator lead, but complete accommodation of its integrated bipolar coil within the right ventricle (RV) can be challenging. Partial coil extension into the right atrium occurred in 6.1% of patients in the LEADR LBBAP trial.1 This challenge may be greater with proximal LBBP, an important consideration because a shorter lead-to-His bundle (HB) distance has been associated with greater improvement in left ventricular ejection fraction.
No takes yet. Share an insight, caveat, or question.
Okubo et al. (2026) studied Left bundle branch pacing. Slack formation technique during proximal left bundle branch pacing with a 4.7-F lumenless defibrillation lead (OmniaSecure) was evaluated. Proximal left bundle branch pacing with a 4.7-F lumenless defibrillation lead faces challenges accommodating the integrated bipolar coil within the right ventricle (6.1% partial RA extension in prior data).