This is the authors' abstract. We don't add key points for this paper.
We have realized an improved resistance calibration setup based on a cryogenic current comparator (CCC). The comparator, with 18 windings and 4647 turns in total, is well-suited for all the necessary comparisons with a 100 Ω standard resistor with the quantum Hall effect and inside the range of standard resistance values from 1 Ω to 1 MΩ. The new state-of-the-art setup is equipped with a low-noise dc SQUID (superconducting quantum interference device) and a digital double current source. Proper damping of the CCC resonance, careful electronics design ('box-in-box') and application of internal wideband feedback to the SQUID sensor improve the dynamic stability considerably and lower the overall measurement time as compared against our previous setup.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Drung et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: