Answers are grounded in this paper and cite their sources.
Reliable force sensing is essential for improving intraoperative safety and precision in minimally invasive surgery (MIS). However, existing forceps-integrated sensors may suffer from electromagnetic interference or position-dependent errors. Here, we develop a 5-mm sensorized surgical forceps integrating cascaded fiber Bragg gratings (FBGs) and dual elliptical flexure hinges for simultaneous clamping-force and position sensing. A force-position estimation method is proposed, where the absolute wavelength-shift ratio of FBG-1 and FBG-2 estimates the clamping position, while their wavelength-shift difference with position-dependent compensation enables force evaluation under concentrated and distributed loading. The forceps achieves average force sensitivities of 44.06 pm/N and 42.31 pm/N under concentrated and distributed loading, respectively, with maximum dynamic errors of 7.11% for force and 2.07% for position. Temperature calibration (25–50 °C) and coupled temperature-force experiments (30–40 °C) verify the effectiveness of the proposed compensation method in reducing temperature-induced errors. Phantom-based suturing and tissue-manipulation tasks further demonstrate the feasibility of simultaneous force-position sensing under surgical interactions.
No takes yet. Share an insight, caveat, or question.
Guo et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: