Key result
Simulated TEER cuts mitral regurgitant fraction by an absolute ~17% but reduces diastolic filling.
Why the study?
Transcatheter edge-to-edge mitral valve repair devices effectively reduce primary mitral regurgitation resulting from chordal rupture, but their impact on global cardiac mechanics and diastolic filling remains difficult to predict.
Does transcatheter edge-to-edge mitral valve repair (TMVr) improve hemodynamic metrics such as regurgitant fraction and left atrial pressure in a computational whole-heart model of chordal rupture-induced primary mitral regurgitation?
Does transcatheter edge-to-edge mitral valve repair (TMVr) improve hemodynamic metrics such as regurgitant fraction and left atrial pressure in a computational whole-heart model of chordal rupture-induced primary mitral regurgitation?
A whole-heart fluid-structure interaction model demonstrates that while transcatheter edge-to-edge mitral valve repair reduces regurgitant fraction and left atrial pressure in chordal rupture-induced mitral regurgitation, it also creates high-velocity inflow jets that reduce diastolic filling.
Purpose Primary mitral regurgitation (MR) resulting from chordal rupture leads to acute volume overload and elevated left atrial pressure. Transcatheter edge-to-edge mitral valve repair (TMVr) devices can effectively reduce MR, but their impact on global cardiac mechanics and diastolic filling remains difficult to predict. This study develops a whole-heart FSI framework for TMVr by extending a previously developed whole-heart model to represent chordal rupture and edge-to-edge repair and to quantify their structural and hemodynamic consequences. Methods A three-dimensional FSI model of the heart derived from cardiac CT imaging was used to simulate three conditions: healthy baseline, chordal rupture-induced MR, and post-TMVr. The immersed finite element–difference method enables fully coupled simulation of blood flow, valve dynamics, and myocardial contraction. Hemodynamic metrics, including regurgitant fraction (RF), left atrial pressure (LAP), stroke volume (SV), stroke work (SW), and pressure–volume relationships, were compared with clinical reference data. Results Chordal rupture produced moderate-to-severe MR (RF = 38%) and doubled left atrial V-wave amplitude, consistent with clinical observations. After TMVr, the central regurgitant jet produced by chordal rupture was eliminated, RF decreased to 21%, and LAP returned to near-baseline levels. However, the double-orifice valve configuration generated dual high-velocity inflow jets and reduced diastolic filling. Conclusion The repair effectively reduces regurgitation and atrial pressure while also influencing diastolic filling and ventricular output. These findings highlight the importance of evaluating both regurgitation reduction and inflow characteristics when assessing TMVr outcomes and demonstrate the potential of patient-specific computational modeling for procedural planning and optimization.
No takes yet. Share an insight, caveat, or question.
Sadipour et al. (2026) studied Primary mitral regurgitation resulting from chordal rupture. Transcatheter edge-to-edge mitral valve repair (TMVr) vs. Chordal rupture-induced mitral regurgitation and healthy baseline was evaluated on Regurgitant fraction (RF), left atrial pressure (LAP), stroke volume (SV), stroke work (SW), and pressure-volume relationships. Simulated transcatheter edge-to-edge mitral valve repair reduced the regurgitant fraction from 38% to 21% and normalized left atrial pressure, but reduced diastolic filling.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: