Key result
Guide-sheath-assisted bailout retrieval enables safe PASCAL disengagement without leaflet injury during intra-procedural SLDA.
Why the study?
Single-leaflet device attachment is a rare but serious complication of mitral valve transcatheter edge-to-edge repair with limited management options after full device release, though intra-procedural recognition before full release may allow a specific bailout strategy.
Does a guide-sheath-assisted retrieval technique enable safe device retrieval and bailout in a patient with intra-procedural single-leaflet device attachment during mitral valve transcatheter edge-to-edge repair using the PASCAL Ace system?
Case Report (n=1)
No
Does a guide-sheath-assisted retrieval technique enable safe device retrieval and bailout in a patient with intra-procedural single-leaflet device attachment during mitral valve transcatheter edge-to-edge repair using the PASCAL Ace system?
This case report describes a novel guide-sheath-assisted retrieval technique that provides a potential bailout strategy for intra-procedural single-leaflet device attachment prior to full implant release during mitral valve transcatheter edge-to-edge repair with the PASCAL system.
May enable bailout for recognized intra-procedural SLDA; leaves open need for prospective validation before broader adoption.
Background Single-leaflet device attachment(SLDA)is a rare but serious complication of mitral valve transcatheter edge-to-edge repair(MTEER). When SLDA occurs after full device release, management options are limited. In contrast, intra-procedural SLDA recognized before full implant release may allow a specific bailout strategy. Case summary A 77-year-old man with severe degenerative mitral regurgitation (MR) underwent MTEER using the PASCAL Ace system because of high surgical risk. Initial device implantation reduced MR from severe to mild. However, after removal of both clasp lines during the release process, SLDA occurred at the posterior leaflet, resulting in recurrent severe MR. Although standard retrieval was no longer feasible, the implant remained connected to the steerable catheter. Under fluoroscopic and transesophageal echocardiographic guidance, a guide sheath was advanced coaxially toward the implant. Controlled insertion of the guide sheath between the clasp and paddle, followed by gradual paddle opening, enabled leaflet disengagement and successful device retrieval without leaflet injury. A second implantation achieved a sufficient reduction in mitral regurgitation. Discussion This case demonstrates a novel bailout technique for intra-procedural SLDA. Successful retrieval requires precise coaxial device alignment, careful catheter manipulation, continuous monitoring of leaflet tension, and meticulous guidance using fluoroscopy and transesophageal echocardiography (TEE). Importantly, prompt recognition of SLDA by TEE is a prerequisite of this technique. In conclusion, the guide-sheath-assisted retrieval technique can represent a potential bailout strategy for intra-procedural SLDA before full implant release during MTEER.
No takes yet. Share an insight, caveat, or question.
Nukariya et al. (2026) conducted a case report in Severe degenerative mitral regurgitation complicated by intra-procedural single-leaflet device attachment (SLDA) during MTEER (n=1). Guide-sheath-assisted bailout retrieval technique using the PASCAL Ace system was evaluated on Successful device retrieval without leaflet injury and subsequent mitral regurgitation reduction. Guide-sheath-assisted bailout retrieval enabled successful PASCAL device disengagement without leaflet injury and reduced severe MR in 1 patient (77-year-old man) with intra-procedural SLDA.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: