Key result
Single-centre S-ICD complication and inappropriate shock rates match the PRAETORIAN trial benchmark.
Why the study?
While large multicentre trials support the safety of subcutaneous implantable cardioverter-defibrillators, real-world clinical outcomes from lower-volume, non-tertiary hospitals are less well characterised.
Does subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation at a single district general hospital result in comparable rates of device-related complications or inappropriate shocks compared to published large trial data (PRAETORIAN S-ICD cohort)?
Cohort (n=81)
No
Does subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation at a single district general hospital result in comparable rates of device-related complications or inappropriate shocks compared to published large trial data (PRAETORIAN S-ICD cohort)?
Real-world S-ICD implantation at a non-tertiary UK district general hospital yielded rates of device-related complications, inappropriate shocks, and appropriate shocks comparable to the landmark PRAETORIAN trial.
Answers are grounded in this paper and cite their sources.
Comparable real-world S-ICD outcomes at a district hospital; extends PRAETORIAN benchmarks to non-tertiary settings but leaves generalizability open.
Background/Introduction Subcutaneous implantable cardioverter-defibrillators (S-ICDs) avoid intravascular leads and reduce long-term lead-related complications when compared with transvenous systems.1,2 While large multicentre trials support their safety, outcomes from lower-volume, non-tertiary hospitals are less well characterised. Purpose To report real-world clinical outcomes following all S-ICD implantations at a single large district general hospital (DGH) in the United Kingdom and provide centre-level benchmarking against published trial data. Methods We retrospectively reviewed consecutive S-ICD implants performed at our centre between 2017–2025. Patients with no follow-up were excluded. The primary composite endpoint included the first occurrence of a device-related complication or an inappropriate shock. Time-to-event analyses used Kaplan–Meier (KM) methods. Crude proportions were compared with the PRAETORIAN S-ICD arm using Fisher’s exact test.3 A secondary time-to-first appropriate shock analysis was performed. Results There were 81 patients in the study cohort (mean age 50 years; 77% male). Reverse-KM median follow-up was 39.4 months. 16 patients (19.8%) met the primary composite endpoint. Components included 10 inappropriate shocks and device-related complications (1 infection, 1 lead fracture, 3 device extractions for device-related issues and 1 generator repositioning). There were 4 deaths, none attributed to the device. The 36-month KM cumulative incidence of the primary composite endpoint was 20.1 % (95 % CI 8.9–29.9), increasing to 26.0 % (95 % CI 12.3–37.5) at 48 months. The event rate was 6.8 per 100 patient-years (95 % CI 3.9–11.1). Compared with the PRAETORIAN S-ICD cohort (68/426 [16%] patients met the primary endpoint), there was no statistically significant difference in crude event proportions (p = 0.42; RR 1.24 [95 % CI 0.76–2.02]).3 13 patients (16.0 %) received at least one appropriate shock during follow-up. The reverse-KM median follow-up for this analysis was 32.5 months, with a 48-month cumulative incidence of 23.3 % (95 % CI 9.1–35.3) and an event rate of 6.0 per 100 patient-years (95 % CI 3.2–10.3). When compared with the PRAETORIAN S-ICD arm, in which 83 of 426 patients (19.5 %) received an appropriate shock, there was again no statistically significant difference (p = 0.42; RR 0.82 [95 % CI 0.49–1.38]).3 Conclusion(s) In this single-centre DGH experience, S-ICD implantation produced event rates and follow-up duration comparable to large trial data. The 4-year cumulative risk of device-related complication or inappropriate shock is clinically notable and underlines the importance of careful patient selection, procedural planning and structured follow-up pathways in DGH practice. Limitations of our study include the retrospective design and modest sample size; further multi-centre DGH data and pooled analyses would improve precision.KM Primary Composite EndpointKM Appropriate Shock
No takes yet. Share an insight, caveat, or question.
Dixon et al. (2026) conducted a cohort in Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation (n=81). Subcutaneous ICD (S-ICD) implantation at a district general hospital vs. PRAETORIAN trial S-ICD cohort was evaluated on First occurrence of a device-related complication or an inappropriate shock (RR 1.24, 95% CI 0.76-2.02, p=0.42).
Synapse has enriched 10 related papers. Consider them for comparative context: