Key result
Electrolytic electroporation achieves up to ~94% transmurality with dose-dependent lesion depth in swine.
Why the study?
Conventional pulsed field ablation requires high voltages and generates a surrounding border zone of reversible electroporation. Electrolytic electroporation is a novel approach designed to convert this reversibly electroporated border zone into durable tissue injury at lower voltages, warranting in vivo feasibility evaluation.
Does electrolytic electroporation (E2) achieve feasible and effective acute cardiac tissue lesion formation across different voltage and pulse protocols in a porcine model?
Does electrolytic electroporation (E2) achieve feasible and effective acute cardiac tissue lesion formation across different voltage and pulse protocols in a porcine model?
This first in vivo porcine demonstration shows that electrolytic electroporation (E2) can achieve dose-dependent, up to 94% transmural myocardial ablation lesions at lower voltages than conventional pulsed field ablation.
E2 supports lower-voltage ablation feasibility in porcine hearts; leaves open human safety, durability, and efficacy trials.
Introduction Pulsed field ablation (PFA) is an emerging ablation technique for the treatment of cardiac arrhythmias. Despite its advantages, PFA requires high voltages to induce irreversible electroporation (IRE), frequently necessitating deep sedation or general anaesthesia. These protocols inherently generate a surrounding border zone of reversible electroporation. Electrolytic electroporation (E2) is a novel PFA approach in which tailored pulse waveforms generate electrolysis products simultaneously with electroporation, thereby converting transient membrane permeabilization into durable tissue injury. By targeting this border zone, E2 represents a fundamentally different ablation strategy with the potential to achieve effective lesion formation at lower applied voltages. Purpose To evaluate the feasibility and acute histopathological effects of E2 for cardiac tissue ablation in a porcine model. Methods Intracardiac ablations were performed using a modified circular pulmonary vein isolation catheter in anesthetized, ventilated pigs (n=3). E2 ablations were delivered to the left atrial appendage, right superior pulmonary vein, superior vena cava, and right atrium. Catheter-tissue contact was confirmed via intracardiac echocardiography. Four tailored pulse waveform protocols were tested across the anatomical locations: 3x750V, 3x1000V, 6x750V, and 6x1000V (Lesions 1–4). Animals were euthanized after 24 hours post-procedure. Macroscopic lesion evaluation was performed using triphenyltetrazolium chloride (TTC) staining, followed by histopathological analysis to determine penetration depth and transmurality. Results E2 applications were successful at all targeted ablation sites. On one occasion, self-terminating atrial fibrillation was observed after pulse delivery; otherwise, no procedural complications were observed. Macroscopic evaluation confirmed distinct lesion formation in all dosage groups. Histopathological analysis showed dose-dependent increases in lesion depth (770 µm, 1795 µm, 1501 µm, and 2495 µm for Lesions 1-4, respectively) and transmurality (42%, 54%, 69%, 94% for Lesions 1-4, respectively). Conclusion E2 is a feasible method for achieving targeted myocardial ablation at lower voltages than conventional PFA. By addressing the reversibly electroporated border zone, this novel approach may enhance lesion durability and reduce the likelihood of conduction recovery. The findings of the study support further investigation to define optimal dosing parameters and to assess chronic lesion durability and electrophysiological outcomes.
No takes yet. Share an insight, caveat, or question.
Benak et al. (2026) studied Cardiac arrhythmias (porcine model for cardiac tissue ablation) (n=3). Electrolytic electroporation (E2) cardiac ablation was evaluated on Feasibility and acute histopathological effects (lesion penetration depth and transmurality). Electrolytic electroporation (E2) achieved dose-dependent myocardial lesion depths (770 to 2495 µm) and transmurality (42% to 94%) in a porcine model.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: