PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
December 6, 2019EP Europace145 citationsOpen Access

Endocardial ventricular pulsed field ablation: a proof-of-concept preclinical evaluation

View Full Paper
JKJacob S. KoruthKKKenji KurokiJIJin Iwasawa

Structured PICO

Does endocardial ventricular pulsed field ablation safely create homogeneous, myocardium-specific lesions in a healthy swine model?

P
Population
4 healthy swine
I
Intervention
Endocardial ventricular pulsed field ablation (PFA) using a prototype 12-Fr deflectable catheter
O
Outcome
Safety and feasibility of ventricular PFA (assessed by electrogram reduction, pacing thresholds, lesion dimensions, and histology)surrogate

Endocardial ventricular pulsed field ablation is feasible and creates homogeneous, myocardium-specific lesions while sparing collateral tissues in a preclinical swine model.

Abstract

AIMS: Pulsed field ablation (PFA) is a novel, non-thermal modality that selectively ablates myocardium with ultra-short electrical impulses while sparing collateral tissues. In a proof-of-concept study, the safety and feasibility of ventricular PFA were assessed using a prototype steerable, endocardial catheter. METHODS AND RESULTS: Under general anaesthesia, the left and right ventricles of four healthy swine were ablated using the 12-Fr deflectable PFA catheter and a deflectable sheath guided by electroanatomic mapping. Using the study catheter, electrograms were recorded for each site and pre-ablation and post-ablation pacing thresholds (at 2.0 ms pulse width) were recorded in two of four animals. After euthanasia at 35.5 days, the hearts were submitted for histology. The PFA applications (n = 39) resulted in significant electrogram reduction without ventricular arrhythmias. In ablation sites where it was measured, the pacing thresholds increased by >16.8 mA in the right ventricle (3 sites) and >16.1 mA in the left ventricle (7 sites), with non-capture at maximum amplitude (20 mA) observable in 8 of 10 sites. Gross measurements, available for 28 of 30 ablation sites, revealed average lesion dimensions to be 6.5 ± 1.7 mm deep by 22.6 ± 4.1 mm wide, with a maximum depth and width of 9.4 mm and 28.6 mm, respectively. In the PFA lesions, fibrous tissue homogeneously replaced myocytes with a narrow zone of surrounding myocytolysis and no overlying thrombus. When present, nerve fascicles and vasculature were preserved within surrounding fibrosis. CONCLUSION: We demonstrate that endocardial PFA can be focally delivered using this prototype catheter to create homogeneous, myocardium-specific lesions.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Koruth et al. (2019) studied this question.

synapsesocial.com/papers/6a063581d13cdeb13e6ff995https://doi.org/10.1093/europace/euz341
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Myocardial Lesion Depth With Circular Electroporation Ablation2012 · 82 citations
  2. 2Minimal coronary artery damage by myocardial electroporation ablation2012 · 115 citations
  3. 3Pulsed Field Ablation for Pulmonary Vein Isolation in Atrial Fibrillation2019 · 678 citations
  4. 4Intracardiac pulsed field ablation: Proof of feasibility in a chronic porcine model2018 · 218 citations
  5. 5Intrapulmonary Vein Ablation Without Stenosis: A Novel Balloon‐Based Direct Current Electroporation Approach2018 · 50 citations