Key result
Neuronal-network pacing with RSA modulation increases stroke volume by ~12 mL in healthy pigs.
Why the study?
Natural heart rate variability driven by neuronal autonomic tone is lost in heart failure, and the left ventricular hemodynamic effects of restoring respiratory sinus arrhythmia (RSA) via a novel 2-lead neuronal-network pacemaker required testing.
Does a novel 2-lead neuronal-network pacemaker incorporating respiratory sinus arrhythmia (RSA) modulation improve left ventricular stroke volume and cardiac output compared to pacing without RSA in a porcine model?
Does a novel 2-lead neuronal-network pacemaker incorporating respiratory sinus arrhythmia (RSA) modulation improve left ventricular stroke volume and cardiac output compared to pacing without RSA in a porcine model?
In a porcine preclinical study, pacing with respiratory sinus arrhythmia modulation via a novel neuronal-network pacemaker improved stroke volume in healthy pigs compared to conventional pacing without heart rate variability.
Answers are grounded in this paper and cite their sources.
RSA-modulated pacing may augment stroke volume in healthy pigs; leaves open translation to human heart failure therapy.
Background Small neural networks known as central pattern generators (CPG) are being increasingly recognized to re-establish physiological heart rate feedback response in pathologies such as heart failure. This approach restores the natural heart rate variability driven by neuronal autonomic tone that is lost in disease or with age. Our novel 2-lead neuronal-network pacemaker is developed to train and restore the physiologic pacing in adaption to breathing, arterial pO2, and blood pressure. In our first-generation pacemaker we were able to induce a paced respiratory sinus arrhythmia (RSA), mimicking a physiologic pattern. (Figure 1) We performed large-animal experiments to test the left ventricular hemodynamics in 2-chamber pacing with and without RSA, in healthy pigs and heart failure model after acute myocardial infarction. Methods A novel 2-chamber pacemaker with a diaphragm respiratory sensor was implanted in healthy pigs (n=5), while conventional single-chamber pacemaker was used in pigs with chronic anterior myocardial infarction (AMI) (n=4) and reduced ejection fraction (44±2,3%). Echocardiographic measurements were performed in both groups at baseline and during variable pacing mimicking the dynamic heart rate sensor at the AMI pigs, with/without preserved RSA at the healthy pigs. Cardiac output (CO) and stroke volume (SV) measurements were performed and compared between the groups. Results Constant pacing of the right ventricle without RSA with variable heart rate sequentially decreased the stroke volume in the AMI pigs from 45.1 (42.8-46.4) mL at the baseline to 42.9 (31.5-46.1) mL at 160bpm. (Figure 2) In the healthy pigs pacing without RSA decreased the stroke volume from 72.35 (69.66-74.42) mL to 56.75 (55.71-59.81) mL. By using our novel 2-lead neuronal-network pacemaker, the increasing RSA modulation during pacing sequentially increased the stroke volume (from 56.75 (55.71–59.81) mL to 63.04 (52.57–65.26) mL at 50% RSA and to 68.68 (68.23–73.2) mL at 100% RSA) at the healthy pigs. Conclusions Pacing with dynamic heart rate variability significantly decreased the stoke volume in AMI pigs with reduced left ventricular ejection fraction, presenting an unmet need to optimize hemodynamics in pacing patterns with dynamic heart rate sensor. In healthy pigs, pacing without RSA decreased the stroke volume, while adding the RSA modulation by our neuronal-network pacemaker improved the stroke volume.
No takes yet. Share an insight, caveat, or question.
Hamzaraj et al. (2026) studied Heart failure / chronic anterior myocardial infarction (porcine model) and healthy controls (n=9). 2-lead neuronal-network pacemaker with respiratory sinus arrhythmia (RSA) modulation vs. Pacing without RSA / conventional single-chamber pacing with variable heart rate was evaluated on Stroke volume (mL) during pacing with vs without respiratory sinus arrhythmia (RSA). Neuronal-network pacing with 100% RSA modulation increased stroke volume in healthy pigs from 56.75 mL (pacing without RSA) to 68.68 mL.