Key result
Flexible microneedle JVP sensor closely matches invasive catheter waveforms.
Why the study?
Jugular venous pressure is a key hemodynamic marker in heart failure, yet current assessment methods remain subjective or invasive, motivating the development of a continuous, non-invasive microneedle-based pressure-sensing platform.
Does a flexible conductive microneedle-based sensor accurately measure jugular venous pressure non-invasively compared to invasive right-heart catheterization in patients undergoing cardiac catheterization?
Does a flexible conductive microneedle-based sensor accurately measure jugular venous pressure non-invasively compared to invasive right-heart catheterization in patients undergoing cardiac catheterization?
Answers are grounded in this paper and cite their sources.
This study demonstrates the first flexible microneedle-based sensor platform capable of continuous, non-invasive jugular venous pressure monitoring with strong correlation to invasive right-heart catheterization measurements.
Introduction Jugular venous pressure (JVP) is a key hemodynamic marker for diagnosis and management of heart failure, yet current assessment methods remain subjective or invasive. A flexible microneedle (MN)-based pressure-sensing platform was developed for continuous, non-invasive JVP monitoring by converting venous pulsation–induced skin deformations into measurable electrical signals. Materials and Methods A soft conductive sensing film composed of 12% ethylene glycol–poly(3,4-ethylenedioxythiophene):polystyrene sulfonate/polyvinyl alcohol (EG–PEDOT:PSS/PVA) was fabricated and mechanically characterized. Polycaprolactone (PCL) MN arrays with varying polyethylene glycol (PEG) concentrations were produced to optimize mechanical coupling. Device performance was evaluated using mechanical testing and electromechanical measurements. A pilot clinical study was conducted in patients undergoing right-heart catheterization following institutional ethics approval and acquisition of informed consent. Results and Discussion The EG–PEDOT:PSS/PVA film exhibited low Young’s modulus (E = 0.0078 ± 0.005 MPa) and high stretchability, enabling sensitive detection of small pressure variations. MN arrays containing 1% PEG demonstrated optimal mechanical properties and significantly enhanced strain transfer. The MN-integrated sensor detected physiologically relevant pressures (2–11 mmHg) with ΔR/R₀ responses of 12–46%, whereas a flat sensor without MNs exhibited only 1–4% variation. In the clinical evaluation (n = 5 patients), stable signal segments showed waveform morphology and temporal features closely matching invasive catheter measurements with a coefficient of determination of R2= 0.7767 across subjects, confirming the effectiveness and clinical relevance of MN-mediated surface sensing. Conclusions This work demonstrates the first microneedle-based platform for continuous, non-invasive JVP monitoring. The technology addresses key limitations of existing clinical methods and offers a novel diagnostic pathway for improved heart failure management.
No takes yet. Share an insight, caveat, or question.
Nachman et al. (2026) studied Heart failure (n=5). Flexible conductive microneedle-based jugular venous pressure sensor vs. Invasive right-heart catheterization measurements was evaluated on Correlation of non-invasive jugular venous pressure waveform with invasive right-heart catheterization measurements (R2 = 0.7767). A flexible microneedle-based JVP sensor showed waveform morphology and temporal features closely matching invasive catheter measurements in 5 patients.
Synapse has enriched 10 related papers. Consider them for comparative context: