Key result
2D-STE yields good inter-vendor agreement for RVFWLS, showing under ~1% bias across software platforms.
Why the study?
Right ventricular and left atrial strain derived from two-dimensional speckle-tracking echocardiography are increasingly used for risk stratification and therapeutic decision-making, but concerns regarding inter-vendor variability and reproducibility continue to limit their routine clinical implementation.
Does two-dimensional speckle-tracking echocardiography provide reproducible and vendor-consistent right ventricular and left atrial strain measurements in human subjects?
Observational (n=62)
Does two-dimensional speckle-tracking echocardiography provide reproducible and vendor-consistent right ventricular and left atrial strain measurements in human subjects?
Right ventricular free-wall and left atrial strain measurements via 2D speckle-tracking echocardiography exhibit high test-retest variability and moderate-to-good inter-vendor agreement, underscoring the need for methodological standardization before routine clinical use.
High test-retest variability cautions against routine clinical adoption of RVFWLS and LA strain; leaves open standardization as a research priority.
Background Right ventricular (RV) and left atrial (LA) strain derived from two-dimensional speckle-tracking echocardiography (2D-STE) are increasingly used for risk stratification and therapeutic decision-making. However, concerns regarding inter-vendor variability and reproducibility continue to limit their routine clinical implementation. Purpose To evaluate inter-vendor agreement, and test–retest reproducibility of RV and LA strain in a standardized, real-world test–retest study design. Methods Sixty-two subjects (50 men; age 56 ± 17 years) underwent echocardiographic acquisitions using six different ultrasound systems. To assess test–retest reproducibility, two image datasets were acquired in each subject by the same experienced echocardiographer under strictly standardized conditions. We performed in total 3348 RV and LA strain analyses from dedicated views using 3 vendor-specific and 6 vendor-neutral 2D-STE software platforms. Results Mid-/full-wall RV free-wall longitudinal strain (RVFWLS) showed good inter-vendor agreement with contemporary software solutions (bias <1strain%), while endocardial RVFWLS demonstrated larger discrepancies (Figure 1). Depending on the software used, test–retest variability for RVFWLS was high to very high (11.3% to 18.3%). Four-chamber LA longitudinal reservoir and contraction strain had a clinically relevant inter-vendor bias (9.3 strain% and 3.7 strain%, respectively, Figure 2). LA contraction strain showed a high test–retest variability (34%). Calculating LA strain as average of both apical views reduced variability significantly (p < 0.05). Conclusions In a strictly controlled, real-world test–retest setting, RV free-wall longitudinal strain showed good to moderate inter-vendor agreement. Measurement variability was 2 to 3 times higher than for left ventricular global longitudinal strain. LA strain, especially during atrial contraction, had only moderate inter-vendor agreement and substantial variability. These findings highlight the need for methodological standardization before widespread implementation of these parameters in routine practice, particularly for left atrial strain.Figure 1Figure 2
No takes yet. Share an insight, caveat, or question.
Balinisteanu et al. (2026) reported an observational. Two-dimensional speckle-tracking echocardiography (2D-STE) for RV and LA strain vs. 3 vendor-specific and 6 vendor-neutral 2D-STE software platforms was evaluated on Inter-vendor agreement and test-retest reproducibility of right ventricular and left atrial strain (p=<0.05). 2D-STE strain showed good inter-vendor agreement for mid-/full-wall RVFWLS (bias <1%) with 11.3-18.3% test-retest variability, while LA contraction strain had 34% variability.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: