Key result
SvO2 and physical activity strongly track cardiac output changes during exercise.
Why the study?
Although methods to measure mixed venous oxygen saturation (SvO2) and physical activity (PhyAc) had been developed, the characteristics of these parameters for total artificial heart control during exercise remained unclear.
Answers are grounded in this paper and cite their sources.
Supports SvO2 and PhyAc sensors for total artificial heart control; leaves open prospective device validation.
Mixed venous oxygen saturation (SvO2) and physical activity (PhyAc) are practical candidates as parameters of total artificial heart (TAH) control because SvO2 can be measured through a transparent blood pump housing with infrared rays and PhyAc can be calculated from signals of an accelerometer used for rate response pacemakers. Although the methods for measurement of the parameters have already been developed, characteristics of these parameters for TAH control, such as during exercise, are still unclear and were examined in this study. SvO2, cardiac output (CO), and PhyAc were measured as parameters. Multi-stage treadmill exercise tests were performed. Difference values (DVs) from the value at the start of exercise showed better correlation than did absolute values. Correlation coefficients between DV in CO and DV in SvO2 and between DV in CO and PhyAc were high at -0.82 and 0.72, while the time constants for the change of SvO2 and CO to the PhyAc change were 26 and 32 sec. Although the correlation coefficient between the CO and SvO2 was higher than that between CO and PhyAc, PhyAc responded more quickly to the speed change compared with the response of SvO2 and CO. It was concluded that SvO2 and PhyAc were useful parameters with different characteristics for TAH control during exercise.
No takes yet. Share an insight, caveat, or question.
OHNO et al. (1997) studied Total artificial heart (TAH) control during exercise. Mixed venous oxygen saturation (SvO2) and physical activity (PhyAc) measurement during multi-stage treadmill exercise was evaluated on Correlation and time constants between difference values (DV) in cardiac output (CO), SvO2, and physical activity (PhyAc) (r = -0.82 (CO vs SvO2); r = 0.72 (CO vs PhyAc)). SvO2 and physical activity (PhyAc) correlated strongly with cardiac output changes during exercise (r = -0.82 and 0.72), with time constants of 26 and 32 sec to PhyAc change.
Synapse has enriched 10 related papers. Consider them for comparative context: