Key result
Absorb BVS meets non-inferiority for 2-year TVF vs XIENCE, despite higher scaffold thrombosis.
Why the study?
This report of the AIDA trial aimed to provide complete two-year outcomes for the primary endpoint of target vessel failure with formal non-inferiority testing and an update on device thrombosis.
Is the Absorb everolimus-eluting bioresorbable vascular scaffold non-inferior to the XIENCE everolimus-eluting metallic stent for preventing target vessel failure at 2 years in patients undergoing routine PCI?
RCT (n=1,845)
Single-blind (patients blinded, operators unblinded)
1:1 centralised web-based system in random block sizes
Yes
Is the Absorb everolimus-eluting bioresorbable vascular scaffold non-inferior to the XIENCE everolimus-eluting metallic stent for preventing target vessel failure at 2 years in patients undergoing routine PCI?
Although the Absorb bioresorbable vascular scaffold formally met non-inferiority criteria for 2-year target vessel failure compared with the XIENCE metallic stent in routine PCI, it carried a significantly higher risk of scaffold thrombosis and target vessel myocardial infarction.
Absorb BVS should not replace metallic stents in routine PCI due to excess thrombosis; challenges bioresorbable scaffold safety assumptions.
AIMS: The aim of this report of the AIDA trial is to provide full two-year outcomes for the primary endpoint of target vessel failure (TVF) and an update on device thrombosis. METHODS AND RESULTS: AIDA was a single-blind, multicentre, investigator-initiated, non-inferiority, randomised (1:1) clinical trial. At complete two-year follow-up, the primary endpoint of TVF had occurred in 100 patients in the Absorb BVS arm versus 90 patients in the XIENCE EES arm (HR 1.12, 95% CI: 0.94-1.49; psuperiority=0.436). Estimated two-year Kaplan-Meier event rates of TVF were 11.0% and 9.9%, respectively (95% CI: -0.9%-3.0%; pnon-inferiority=0.003). Definite or probable device thrombosis at two years occurred in 30 patients in the Absorb BVS arm and in eight patients in the XIENCE EES arm. Kaplan-Meier estimates of device thrombosis were 3.3% in the Absorb BVS arm and 0.9% in the XIENCE EES arm (HR 5.22, 95% CI: 2.00-13.59; p<0.001). CONCLUSIONS: AIDA formally met its criterion for non-inferiority of Absorb BVS versus XIENCE EES in terms of the combined endpoint of TVF. The Absorb BVS, however, was associated with higher rates of scaffold thrombosis and target vessel myocardial infarction at complete two-year follow-up.
No takes yet. Share an insight, caveat, or question.
Kraak et al. (2018) conducted an RCT in Coronary artery disease undergoing routine PCI (n=1,845). Absorb everolimus-eluting bioresorbable vascular scaffold (Absorb BVS) vs. XIENCE everolimus-eluting metallic stent (XIENCE EES) was evaluated on Target vessel failure (composite of cardiac death, myocardial infarction, and target vessel revascularisation) at 2 years (HR 1.12, 95% CI 0.94-1.49, p=0.003 for non-inferiority (0.436 for superiority)). Absorb BVS was non-inferior to XIENCE EES for 2-year target vessel failure (11.0% vs 9.9%; HR 1.12; p_non-inferiority=0.003), despite higher scaffold thrombosis (3.3% vs 0.9%, p<0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: