Why the study?
Small indexed estimated orifice area predicts poor outcomes with surgical bioprosthetic valves, whereas balloon-expandable and self-expanding valves show more variable results, prompting evaluation of valve geometry effects on gradient overestimation and alternative body size adjustments beyond body surface area.
Does valve geometry affect the overestimation of Doppler gradients across different prosthetic aortic valve types, and are alternative body size measures superior to body surface area for predicting stroke volume?
Does valve geometry affect the overestimation of Doppler gradients across different prosthetic aortic valve types, and are alternative body size measures superior to body surface area for predicting stroke volume?
Prosthetic aortic valve geometry significantly impacts hemodynamic gradient measurements, and adjusting for left ventricular outflow tract area rather than body surface area may allow more accurate evaluation of patient-prosthesis mismatch.
Patient-prosthesis mismatch (PPM) is defined by the estimated orifice area divided by body surface area (BSA), or iEOA. Small iEOA predicts poor outcomes with surgical bioprosthetic valves (SBV), but balloon-expandable valves (BEV) and self-expanding valves (SEV) have more variable results. Assess effect of geometry of the valve on overestimation of gradients and methods for adjusting for body size other than BSA. Computational fluid dynamic simulations of SBV, BEV and SEV, with 25 different variations of anatomic and valve sizes for each valve type determined peak stenosis velocity and peak and total valve gradients. The relationship between stroke volume and body size measures was analyzed using data from the Atherosclerosis Risk In Communities study. Doppler peak velocity correctly estimated the total gradient for SBV (slope = 1.01) but overestimated the gradient for BEV (slope = 1.19) and SEV (slope = 1.42). The gradient across the valves varied even with identical dimensions, with SBV being highest, BEV lower, and SEV lowest. BSA had a fair correlation with stroke volume (r = 0.6030), but lean body mass (r = 0.6494) and fat-free mass (r = 0.6631) were better and left ventricular outflow tract (LVOT) area was best (r = 0.7570). Geometry of the valve has a major impact on hemodynamics with very different results with identical dimensions. LVOT area was better way to adjust for body size than BSA. Two measures based on LVOT velocity and area are proposed as an alternative to implement these results. If these findings are confirmed with in-vitro and in-vivo studies, the approach could allow more accurate prediction and evaluation of PPM.
No takes yet. Share an insight, caveat, or question.
David T. Linker (2026) studied Patient-prosthesis mismatch in prosthetic aortic valves. Balloon-expandable valves (BEV) and self-expanding valves (SEV); LVOT area for body size adjustment vs. Surgical bioprosthetic valves (SBV); Body surface area (BSA) was evaluated on Estimation of total valve gradient via Doppler peak velocity and correlation of body size measures with stroke volume (r = 0.7570 for LVOT area vs r = 0.6030 for BSA; slope = 1.19 (BEV) and 1.42 (SEV) vs 1.01 (SBV)).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: