Key result
PPG smartband screening detects incident post-stroke AF with ~89% NPV but low ~32% PPV.
Why the study?
Detecting AF in stroke survivors is critical for preventing recurrent stroke, yet the diagnostic accuracy of photoplethysmography-based smartwatches in this high-risk population remained uncertain.
Does a photoplethysmography-based Huawei smartband worn for one month accurately detect incident and known atrial fibrillation in recent ischaemic stroke survivors?
Observational (n=491)
Yes
Does a photoplethysmography-based Huawei smartband worn for one month accurately detect incident and known atrial fibrillation in recent ischaemic stroke survivors?
In recent ischaemic stroke survivors, one-month monitoring with a PPG-based smartband demonstrated a high negative predictive value (89%) for ruling out atrial fibrillation, though a low positive predictive value (32%) necessitates confirmatory ECG monitoring.
Answers are grounded in this paper and cite their sources.
Captured external expert commentary on this paper, strongest first. Original sources are linked where available.
“We are delighted by this collaboration with Brainomix, which enhances our growing research portfolio into stroke and atrial fibrillation research, and would help inform clinical practice and improve our care and management of these patients.”
May support AF screening post-stroke; leaves open validation against ECG before routine adoption.
Background Detecting atrial fibrillation (AF) in stroke survivors is critical for preventing recurrent stroke. However, the diagnostic accuracy of photoplethysmography (PPG)-based smartwatches in this high-risk population remains uncertain. The Liverpool-Huawei Stroke Study evaluated the performance of Huawei smartbands for AF detection following an ischaemic stroke. Purpose 1. To determine the clinical effectiveness of Huawei smartband by assessing the negative predictive value (NPV) and positive predictive value (PPV) for identifying incident and known AF in a post-stroke population. 2. To identify clinical predictors associated with AF detection. Methods Between June 2022 and June 2024, patients hospitalised with a recent ischaemic stroke were recruited from 8 National Health Service Trusts. Participants wore a Huawei smartband paired with an Android smartphone for one month. The device automatically screened for AF at regular intervals, transmitting data via the Huawei Health App. Following alerts generated by the device, clinical evaluation with ECG and 5-14 days Holter or ECG patch monitoring were used to confirm AF diagnosis although these were not done concurrent with the alert. Results 491 patients (median age: 62 [QR 54, 72]; 35% female) were recruited (Table). During the one-month monitoring period, 192 (39.1%) individual AF alerts were received, 15 (7.8%) were confirmed as incident AF, 114 (59.4%) were confirmed AF-negative, and 39 (7.9%) identified as known AF. (Figure). The resulting NPV and PPV were 0.89 (0.85, 0.93) and 0.32 (95% CI: [0.25, 0.40]), respectively. Multivariable logistic regression indicated that age (OR: 1.06, 95%CI [1.03, 1.11]), current smoker (0.35 [0.10, 0.96]) and systolic blood pressure (0.98 [0.96, 0.99]) were independently associated with AF detection (Figure). Conclusions The Huawei smartband revealed incident AF in 15 (7.8%) patients post-stroke with a high NPV (89%), but a low PPV (32.1%). This may reflect that ECG monitoring was not done concurrent with the alert and ectopic beats (common in this population) were flagged as an alert due to irregular PPG rhythm; however, the absence of an alert likely reflects the absence of AF post-stroke. Baseline age, smoking status, and systolic blood pressure were independent predictors for AF detection.
No takes yet. Share an insight, caveat, or question.
Lam et al. (2026) conducted an observational in Atrial fibrillation in ischaemic stroke survivors (n=491). Photoplethysmography (PPG)-based Huawei smartband paired with an Android smartphone vs. Clinical evaluation with ECG and 5-14 days Holter or ECG patch monitoring was evaluated on Diagnostic accuracy (NPV and PPV) for identifying incident and known atrial fibrillation (NPV 0.89; PPV 0.32, 95% CI NPV: 0.85-0.93; PPV: 0.25-0.40). PPG-based Huawei smartband monitoring post-stroke detected incident AF in 7.8% of alert patients with a high NPV of 0.89 and a low PPV of 0.32.
Synapse has enriched 10 related papers. Consider them for comparative context: