Key result
Clover tricuspid edge-to-edge repair acutely reduces septal-lateral annular diameter by ~3 mm.
Why the study?
Transcatheter edge-to-edge repair reduces tricuspid regurgitation with some patients showing an additional annuloplasty effect, motivating comparison of the Clover versus Zipping clip configurations on tricuspid annular geometry and dynamics.
Does the Clover configuration compared to the Zipping configuration improve acute tricuspid regurgitation reduction and tricuspid annular septal-lateral diameter reduction in an ovine model of functional tricuspid regurgitation?
Does the Clover configuration compared to the Zipping configuration improve acute tricuspid regurgitation reduction and tricuspid annular septal-lateral diameter reduction in an ovine model of functional tricuspid regurgitation?
In an ovine model of functional tricuspid regurgitation, both Clover and Zipping edge-to-edge repair techniques reduced tricuspid regurgitation, but only the Clover configuration produced a significant acute annuloplasty effect by reducing septal-lateral annular diameter while preserving annular contraction.
Answers are grounded in this paper and cite their sources.
Clover configuration may confer additional annular reduction in functional TR; hypothesis-generating for technique selection pending human data.
Introduction: Transcatheter edge-to-edge repair effectively reduces tricuspid regurgitation (TR), with some patients showing an additional 'annuloplasty effect'. Methods: Ten adult sheep underwent pulmonary artery banding to induce right ventricular failure and functional TR. Six weeks later, sonomicrometry crystals were implanted around the tricuspid annulus (TA). Two Abbott XTW clips were then placed using either a Zipping (n=5) or Clover (n=5) configuration. Echocardiographic and sonomicrometry data were collected before and after clip implantation to assess TA geometry and function. Results: The model produced right ventricular dilation, dysfunction, and moderate-to-severe functional TR. In all animals, leaflet approximation with two clips significantly reduced TR severity (3.6+/-0.5 to 1.3+/-1.1; p=0.01) and septal-lateral annular diameter (29.2+/-2.2 to 26.7+/-2.9 mm; p=0.01). However, a significant reduction in septal-lateral annular diameter occurred only with the Clover technique (30+/-2 to 27+/-1 mm; p=0.001). Conclusions: Both techniques reduced TR, but only the Clover configuration produced an acute annuloplasty effect by selectively decreasing septal-lateral annular diameter.
No takes yet. Share an insight, caveat, or question.
Piekarska et al. (2026) studied Functional tricuspid regurgitation and right ventricular failure (n=10). Tricuspid edge-to-edge repair (TEER) using Clover configuration vs. Pre-clipping baseline (and comparison with Zipping configuration using two clips between septal and anterior leaflets) was evaluated on Tricuspid annular septal-lateral diameter (mm) with Clover technique (MD -3.0 mm, p=0.001). Clover configuration tricuspid edge-to-edge repair acutely reduced tricuspid annular septal-lateral diameter from 30±1.5 to 27±0.8 mm and TR grade from 3.8±0.4 to 1.1±0.5.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: