Answers are grounded in this paper and cite their sources.
Bulk analytical methods are extensively used to identify molecular cargo in extracellular vesicles (EVs), but they cannot distinguish nanovesicles from co-isolated non-vesicular particles or resolve discrete EV subpopulations. Single-particle interferometric reflectance imaging coupled with immunofluorescence (SP-IRIS/IF) has emerged as a platform to simultaneously resolve particle size, count, and cargo at single-EV resolution. However, differentiating authentic EVs from non-vesicular signals requires rigorous methodological controls that remain widely underreported in standard workflows. In this study, we analyzed human neural secretomes and serum EVs across different SP-IRIS/IF setups to evaluate their analytical specificity. By using detergent-mediated lysis controls, we found a critical technical limitation in standard single-particle analysis: the persistence of both interferometric and fluorescent signals after detergent treatment in a subset of presumed single-EV events that were non-vesicular artifacts. These findings indicate that lacking appropriate controls on single-EV platforms risks data misinterpretation, underscoring the indispensable implementation of detergent-mediated lysis for single-particle analysis in SP-IRIS/IF platforms.
No takes yet. Share an insight, caveat, or question.
Guendulain et al. (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: