Key result
Emergent valve-in-valve implantation resolves severe intravalvular AR caused by pigtail catheter entrapment during TAVI.
Why the study?
Acute prosthetic leaflet dysfunction after TAVI is rare and potentially life-threatening, and catheter-induced leaflet trauma during post-dilatation is less well recognised.
Case Report (n=1)
No
This case report highlights pigtail catheter entrapment during TAVI post-dilatation as a preventable cause of acute prosthetic leaflet damage and severe intravalvular regurgitation, managed successfully with valve-in-valve implantation and chimney stenting.
Background Acute prosthetic leaflet dysfunction after transcatheter aortic valve implantation (TAVI) is rare but potentially life-threatening. Catheter-induced leaflet trauma during post-dilatation is less well recognised. Case summary An 88-year-old woman with severe symptomatic aortic stenosis and high-risk aortic root anatomy—including a narrow sinus of Valsalva, low left coronary artery (LCA) height, and a small, calcified sinotubular junction—underwent elective TAVI with a 23-mm Sapien 3 Ultra RESILIA valve deployed with a 2-mL underfill under prophylactic LCA protection. Initial transthoracic echocardiography and aortography showed trivial paravalvular leak and preserved LCA flow. During post-dilatation, the distal balloon tip and Safari guidewire snagged the pigtail catheter, pulling it across the prosthesis and into the left ventricle. Unrecognised in a crowded fluoroscopic field, the catheter became trapped between the expanding balloon and the valve frame, causing focal leaflet injury and severe intravalvular aortic regurgitation with haemodynamic collapse. Emergent valve-in-valve implantation with a 20-mm Sapien 3 Ultra RESILIA under bilateral coronary protection resolved the regurgitation, but displaced leaflet tissue caused LCA ostial compromise requiring chimney stenting. The patient recovered uneventfully and remained stable at 30-day follow-up. Discussion This case illustrates pigtail catheter entrapment during post-dilatation as a preventable cause of acute prosthetic leaflet damage and severe intravalvular regurgitation. It underscores the need for meticulous catheter management, critical reassessment of post-dilatation in older patients with trivial leak, and individualised valve platform and size selection in small, anatomically constrained aortic roots.
No takes yet. Share an insight, caveat, or question.
Tokuyama et al. (2026) conducted a case report in Severe symptomatic aortic stenosis with acute prosthetic leaflet dysfunction and severe intravalvular aortic regurgitation following TAVI (n=1). Emergent valve-in-valve TAVI (20-mm Sapien 3 Ultra RESILIA) with LCA chimney stenting after post-dilatation pigtail catheter entrapment was evaluated on Resolution of severe intravalvular aortic regurgitation and clinical stability at 30 days. Pigtail catheter entrapment during post-dilatation of a 23-mm Sapien 3 Ultra RESILIA valve caused severe intravalvular AR in an 88-year-old woman, resolved via a 20-mm valve-in-valve and LCA stenting.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: