Key result
CT-guided transcatheter closure achieves ascending aortic pseudoaneurysm exclusion without coronary obstruction.
Case Report (n=1)
Captured external expert commentary on this paper, strongest first. Original sources are linked where available.
“CT sizing defines the pseudoaneurysm neck but not where the expanded retention disc will sit, which is the critical relationship when the neck lies near a coronary ostium.”
Preprocedural geometric projection of the expanded occluder footprint onto 3D CT reconstructions can guide device selection, working projection, and coronary protection during transcatheter closure of ascending aortic pseudoaneurysms near coronary ostia.
Ascending aortic pseudoaneurysm is a rare late complication of aortic surgery. Transcatheter closure is an option for patients at prohibitive surgical risk, but when the neck lies near a coronary ostium, device sizing alone does not predict the final position of the expanded disc. An 82-year-old woman with prior type A dissection repair and subsequent root replacement developed an enlarging ascending aortic pseudoaneurysm with a 15.8 × 13.1 mm neck adjacent to the right coronary ostium. Before intervention, the anticipated footprint of a 12-mm septal occluder was projected geometrically onto 3-dimensional computed tomography reconstruction using the intended working projection; the ostium remained just outside the predicted disc margin. The coronary artery was wired prophylactically, the device seated as predicted, and angiography and echocardiography confirmed sac exclusion with an unobstructed vessel. Projecting the expanded occluder footprint onto computed tomography settled device selection, working projection, and coronary protection preprocedurally rather than by intraprocedural assessment.
No takes yet. Share an insight, caveat, or question.
Saeed et al. (2026) conducted a case report in Ascending aortic pseudoaneurysm near the right coronary ostium (n=1). Transcatheter closure with a 12-mm septal occluder guided by geometric 3D CT footprint projection and prophylactic coronary wiring was evaluated on Pseudoaneurysm sac exclusion with an unobstructed right coronary artery. Transcatheter closure of a 15.8 × 13.1 mm neck ascending aortic pseudoaneurysm using a 12-mm septal occluder guided by 3D CT footprint projection achieved sac exclusion without coronary obstruction.