Key result
Freedom Force long coronary stenting yields high success but carries a ~38% 6-month restenosis rate.
Why the study?
The study describes the preliminary experience with stent-supported coronary angioplasty in long narrowings using the Freedom Force long stent with an innovative design.
What are the acute and 6-month clinical and angiographic outcomes of using the Freedom Force long coronary stent in patients with coronary narrowings >20 mm?
Observational (n=27)
What are the acute and 6-month clinical and angiographic outcomes of using the Freedom Force long coronary stent in patients with coronary narrowings >20 mm?
Preliminary experience suggests the flexible Freedom Force long stent enables high acute procedural success in long, tortuous coronary lesions using 1 or 2 stents, though 6-month angiographic restenosis remains at 38%.
Answers are grounded in this paper and cite their sources.
High procedural success with no MACE supports feasibility in long lesions; leaves open whether restenosis rates improve versus modern stents.
This report describes our preliminary experience with coronary stent-supported angioplasty in long narrowings using a long stent with an innovative design. Twenty-seven consecutive patients with target lesions >20 mm in length had a stenting procedure using the Freedom Force long coronary stent (Global Therapeutics, Inc., Broomfield, CO). Target lesion length ranged from 20.7-57.5 mm (mean, 27.66+/-9.41 mm). A total of 35 stents was implanted with a mean stented length of 36.26+/-12.36 mm. The stenting procedure was successful in all patients. Single long stent implantation was performed in 19 patients, while 8 patients had double stent implantation. No major cardiac adverse events occurred during hospital stay. The restenosis rate at the 6-mo angiographic follow-up was 38% (follow-up rate, 96%). During follow-up, no major cardiac events such as death, myocardial infarction, or coronary artery surgery occurred, while 3 patients (11%), all with recurrent angina and angiographic restenosis, underwent repeat coronary angioplasty. Potential advantages of this innovative stent in long narrowings relate to its high flexibility in passing through long tortuous diseased segments, and in treating long lesions using only 1 or 2 stents.
No takes yet. Share an insight, caveat, or question.
Antoniucci et al. (1998) conducted an observational in Long coronary artery narrowings (>20 mm) (n=27). Freedom Force long coronary stent was evaluated on Angiographic restenosis at 6 months. Freedom Force long coronary stenting achieved 100% procedural success in lesions >20 mm, with a 38% 6-month restenosis rate and no major cardiac events.
Synapse has enriched 10 related papers. Consider them for comparative context: