High-purity germanium detectors (impurity concentration ≤1010cm-3) are widely used in precision radiation detection due to their extremely low noise and excellent energy resolution. However, their complex fabrication processes and high cost severely limit large-scale applications. To address this, we have developed a detector based on an a-Ge:H/Al2O3 dual passivation layer, which adopts a low-cost 6N germanium substrate with a doping concentration of 1012cm-3and a thickness of 500 μm, featuring a 5 × 5 mm2 active region surrounded by the guard rings, aiming to break through the technical bottleneck that low-purity germanium struggles to achieve efficient radiation detection by suppressing surface and contact defects. At 90 K, the detector interface state density is significantly reduced from 2.8×1012eV-1cm-2to 3.6×1011eV-1cm-2, and the reverse leakage current (@50 V) is drastically decreased from approximately 200 pA to 20 pA. The detector with the a-Ge:H/Al2O3dual passivation layer achieves an energy resolution of 376 keV (6.85%) under for241Am source (5.486 MeV), exhibiting a well-defined spectrum peak. In contrast, the detector without this dual passivation layer shows no distinct peak. This study lays a foundation for the application of low-purity germanium in portable radiation monitoring and integrated detection arrays, and provides key passivation technology support for the development of thick detectors.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: