Answers are grounded in this paper and cite their sources.
In minimally invasive surgeries (MIS), continuum manipulators made from soft materials make it possible to reach deep-seated lesions with improved contact safety. However, their practical applications are often constrained by nonlinear actuation behavior and limited load-bearing capacity. In this article, a cable-actuated soft manipulator with variable stiffness is designed based on fibrous materials. The manipulator’s skeleton features a hybrid braided fiber structure, offering flexibility while maintaining profile stability, which helps mitigate the nonlinear effects of cable actuation. A fiber jamming mechanism generated by negative pressure allows a rapid stiffness transition between rigid and flexible states. Both the kinematics and stiffness are modeled and validated, demonstrating a linear kinematical response with an average positioning error of 1.94% and a sixfold increase in stiffness through jamming. Capabilities of path tracking and force output are also experimentally evaluated. Finally, a prototype is fabricated and put inin vitromanipulation tests, confirming its potential for MIS applications.
No takes yet. Share an insight, caveat, or question.
Shang et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: