This is the authors' abstract. We don't add key points for this paper.
In this work, we fabricate and study a sensitive polymer-based optical sensor for visible range detection in water environment. The studied transducer is a micro-resonator relying on the interaction between the evanescent wave and the aqueous medium to measure the medium refractive index change. The micro-resonator consists of a waveguide with PMMA polymer as the core layer and porous silica as the lower cladding layer to significantly enhance the evanescent field ratio and sensitivity. The first results show a very good resonator quality factor of around 2.10⁴and a sensitivity of 255nm/RIU at 760nm for aqueous solutions with glucose at different concentrations. These state-of-the-art results obtained in the visible range demonstrate that a polymer-based sensor with a porous cladding appears as a promising platform for evanescent field-based sensing in the visible range without any surface chemical functionalization in an aqueous medium.