On unseen patients, Raman spectroscopy separates breast cancer from normal tissue at about 91% accuracy, but separates DCIS, ILC and IDC at only about 59%.We re-evaluated machine-learning classifiers on the public Imperial College ex vivo breast Raman dataset (641 spectra, 80 samples, 71 donors) using donor-disjoint validation, with preprocessing and classifier settings re-implemented from the deposited pipeline.The number of principal components was chosen by nested cross-validation inside training donors only.Over 20 random seeds, a support vector machine reached 90.8% accuracy (seed SD 0.6) for cancer versus normal and 59.3% (SD 1.3) for four-class subtype classification, with ILC sensitivity of 17.9%.Published figures of 97.7% and 91.0% were not reproduced under donor-disjoint validation; a literal reading of the cross-validation step in the deposited MATLAB scripts reproduces them to within about 3 points.Split-conformal prediction calibrated on separate donors gave 89.1% coverage at a 90% target for cancer versus normal, but only 83.8% coverage and 31% single-label predictions for four classes.In simulations calibrated to the real variance structure, spectrum-level validation overestimated accuracy on new donors by 12 to 23 points, while donor-level validation was nearly unbiased.We recommend donor-level splits, nested model selection and calibrated abstention for Raman tissue classifiers.