Physical performance tests such as the 30-s sit-to-stand (30-s STS) test, Timed Up and Go (TUG), standing-balance tasks, the five-times sit-to-stand (5xSTS) test, and gait speed are widely used to quantify functional capacity. This article describes a reproducible validation workflow for a two-dimensional (2D) red-green-blue (RGB) camera-based markerless motion capture (MMC) application. The workflow combines standardized test administration, simultaneous RGB and depth-camera acquisition, independent clinician scoring, an operational tracking-quality rule, independent criterion annotation, test-retest assessment, second-clinician rescoring, and robustness testing under altered acquisition conditions. In a verified pilot implementation involving 15 healthy adults and seven physical-performance tests, 203 of 210 initial planned standard-condition acquisitions before technical replacement were analyzable (96.7%). Application-to-criterion intraclass correlation coefficients (ICCs) ranged from 0.781 to 0.972 across tests; the lower value for side-by-side standing occurred in the setting of restricted between-participant range and ceiling behavior. Stress-condition analyzability was 43/54 (79.6%). These pilot results characterize feasibility, agreement, repeatability, and failure modes; they are not interpreted using a retrospectively imposed binary validation threshold.
No takes yet. Share an insight, caveat, or question.
Yue et al. (2026) studied this question.