Key result
LBBAP reimplantation linked to ~17-minute shorter procedure time vs epicardial pacing after CRT-D extraction.
Why the study?
CRT-D reimplantation after transvenous lead extraction (TLE) can be challenging due to fibrotic remnants and absent target veins. Although left bundle branch area pacing (LBBAP) is emerging as an alternative to conventional CRT, data regarding its feasibility and safety after TLE were lacking.
Does left bundle branch area pacing (LBBAP) improve procedural time, QRS duration, and electrical lead parameters compared to conventional CRT through epicardial pacing in patients undergoing CRT-D transvenous lead extraction and reimplantation?
Cohort (n=26)
No
Does left bundle branch area pacing (LBBAP) improve procedural time, QRS duration, and electrical lead parameters compared to conventional CRT through epicardial pacing in patients undergoing CRT-D transvenous lead extraction and reimplantation?
In patients requiring CRT-D reimplantation after transvenous lead extraction, left bundle branch area pacing is a feasible alternative to epicardial pacing that offers shorter procedure and fluoroscopy times, narrower QRS duration, and lower pacing thresholds.
Answers are grounded in this paper and cite their sources.
LBBAP was associated with shorter procedures and better parameters after extraction; leaves open outcome benefits versus epicardial pacing in trials.
Background Cardiac resynchronization therapy – defibrillator (CRT-D) reimplantation after transvenous lead extraction (TLE) can be challenging due to fibrotic remnants and absence of target veins. Left bundle branch area pacing (LBBAP) is emerging as an alternative to conventional CRT, but data is lacking about its feasibility after TLE. Purpose To explore the feasibility and safety of LBBAP as alternative to conventional CRT though epicardial pacing (EP) in patients undergoing CRT-D TLE. Methods We prospectively enrolled all patients admitted for complete CRT-D extraction and replacement from July 2022 to September 2023. At the moment of reimplantation, the choice between EP and LBBAP was made on the basis of posterolateral CS branches viability at post-extraction CS venography. Procedural success was defined as effective stimulation and absence of major complications while the primary endpoints were procedural time, QRS complex duration and electrical lead parameters. Patients undergoing LBBAP and EP reimplantation were compared. Results Twenty-six patients were enrolled (mean age 68.3 ± 7.2 years, p = 0.538; 12% females, p = 0.859). The most frequent cause of device extraction was pocket infection (58%, p = 0.773). Seventeen (65%, p = 0.783) patients had estimated high risk of procedural severe complications measured by EROS score. Overall, 80 leads were extracted. Mechanical extractors and snares were deemed necessary in 35 cases (43%, p = 0.917) and in 2 cases, respectively. Almost all patients (n=25, 96%) underwent right-sided reimplantation. Procedural success was achieved in 24 cases (92%). Compared to standard CRT through EP, LBBAP reimplantation accounted for a faster procedure (109.5 ± 34.2 vs. 126.5 ± 46.8 min; p = <0.001), less radiologic exposure (877.0 ± 234.6 vs. 1332.0 ± 515.1 sec, p <0.001) and better ventricular pacing thresholds (0.7± 0.1 vs. 1.8± 0.6 V, p <0.001). Despite a similar pre-extraction paced QRS duration (127.4 ± 13.8 vs. 133.0 ± 14.4 msec; p = 0.610), LBBAP reported significant improvement in final QRS duration (105.0 ± 13.1 vs. 127.1 ± 10.4 msec, p < 0.001) with major reduction in QRS duration (- 10.9 ± 7.5 vs - 6.8 ± 10.1, p < 0.001). Conclusion After a CRT TLE procedure, LBBAP reimplantation accounted for lower procedure duration, lower radiologic exposure, narrower QRS and lower ventricular pacing thresholds, demonstrating to be a valuable alternative to EP.
No takes yet. Share an insight, caveat, or question.
Baroni et al. (2024) conducted a cohort in Patients undergoing complete CRT-D transvenous lead extraction and replacement (n=26). Left bundle branch area pacing (LBBAP) vs. Conventional CRT through epicardial pacing (EP) was evaluated on Procedural time, QRS complex duration, and electrical lead parameters (p=<0.001). LBBAP reimplantation after CRT-D extraction had shorter procedure time, narrower QRS, and lower thresholds vs epicardial pacing.
Synapse has enriched 10 related papers. Consider them for comparative context: