Answers are grounded in this paper and cite their sources.
This work presents methods enabling real-time multi-species direct absorption spectroscopic fitting at kilohertz iteration rates with high accuracy of spectral line shapes on embedded systems. Precomputation of multi-dimensional polynomials and effective pressures allows high accuracy approximation of HITRAN Hartmann-Tran absorption coefficients with full matrix broadening and low residual error, replacing computationally slow line-by-line simulations and inaccurate fixed-basis functions. A novel Levenberg--Marquardt solver augmented with Taguchi orthogonal array exploration accelerates fitting convergence and escapes local minima. Together, these methods enable simulated implementations that fit 1000-point spectra with simultaneous etalon fringe, temperature, and pressure estimation at 1~kHz using 0.8~W simulated computational power for a single laser system. This work further simulates a simultaneous dual-laser single-detector architecture exploiting co-prime spectral lengths, fitting six concentration parameters at 0.82~kHz using 1.6~W simulated computational power. Mid-infrared TDLAS simulations targeting CH4, H2O, C2H6, CO2, and N2O provide an approach for possible atmospheric sensing deployments and chemical kinetics studies.
No takes yet. Share an insight, caveat, or question.
Stephen So (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: