Key result
Simultaneous chimney stenting with Evolut Pro+ TAVR preserves coronary perfusion and access in human heart micro-CT.
Why the study?
Coronary artery occlusion is a potentially devastating complication of TAVR, and multimodal imaging of simultaneous chimney stenting and TAVR in a reanimated human heart provides data in real human anatomy to educate physicians and inform future TAVR technologies.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
Supports feasibility in reanimated heart model; leaves open in vivo safety, efficacy, and clinical outcome trials.
Transcatheter aortic valve replacement (TAVR) has become a popular treatment option for severe aortic stenosis for patients with a high risk for mortality with surgical aortic valve replacement (SAVR). Coronary artery occlusion (CAO) following the implantation of the device is a potential and sometimes devastating complication of this procedure, that provokes a sudden deterioration of hemodynamic status followed by cardiogenic shock and electrical instability. With patients that present a high risk for coronary obstruction, coronary protection with a chimney stenting technique is an effective strategy that can ensure coronary perfusion during TAVR in case of acute CAO. Utilizing Visible Heart® methodologies, a human heart was reanimated. A chimney stenting technique was implemented simultaneously with the deployment of a Medtronic Evolut™ Pro+ valve (Medtronic PLC; Minneapolis, MN, USA). The entire procedure was recorded utilizing endoscopic cameras, fluoroscopy, optical coherence tomography, and echocardiography. In addition to these procedural visualizations, post-procedural micro-computed tomography (micro-CT) was conducted to provide post-implantation imaging with approximately 60-micron resolution. Utilizing these imaging modalities in a reanimated human heart allows for the unique opportunity to collect data for TAVR procedures in real human anatomies for the subsequent educational uses by the physicians treating aortic valvular disease and/or the designers of future TAVR technologies and procedures.
No takes yet. Share an insight, caveat, or question.
Bielecki et al. (2022) conducted a case report in High risk for coronary artery occlusion during TAVR due to low proximal positioning of the right coronary ostium (n=1). Simultaneous chimney stenting (4 mm x 22 mm Onyx stent) and TAVR (26 mm Medtronic Evolut Pro+ valve) was evaluated on Coronary ostium flow access, perfusion, and prosthesis placement assessed via multimodal imaging and micro-CT (60-micron resolution). Simultaneous chimney stenting (4x22 mm Onyx) with a 26 mm Evolut Pro+ TAVR preserved right coronary perfusion and left coronary access in a reanimated human heart, verified by 60-micron micro-CT.
Synapse has enriched 10 related papers. Consider them for comparative context: