Key result
Microcatheter manipulation successfully reseats a displaced WEB device, restoring branch flow and avoiding rescue stenting.
Why the study?
Post-detachment Woven EndoBridge displacement is typically managed with rescue stenting, which necessitates unplanned long-term dual antiplatelet therapy, motivating this report of an alternative bailout using microcatheter manipulation alone.
Does mechanical manipulation via a microguidewire and microcatheter ('pushing technique') restore parent vessel flow and reseat the device without requiring rescue stenting in a patient with post-detachment Woven EndoBridge displacement?
Case Report (n=1)
No
Does mechanical manipulation via a microguidewire and microcatheter ('pushing technique') restore parent vessel flow and reseat the device without requiring rescue stenting in a patient with post-detachment Woven EndoBridge displacement?
This case demonstrates that post-detachment displacement of a Woven EndoBridge device during intracranial aneurysm embolization can be safely managed using a microcatheter 'pushing technique' to restore vessel patency, thereby avoiding rescue stenting and mandatory long-term dual antiplatelet therapy.
Answers are grounded in this paper and cite their sources.
Should not yet change bailout strategies for displaced WEB devices; leaves open the reliability of microcatheter pushing in larger series.
OBJECTIVE: To report a case of post-detachment Woven EndoBridge (WEB) displacement successfully managed through microcatheter manipulation alone, avoiding unplanned stenting, which necessitates long-term dual antiplatelet therapy. CASE PRESENTATION: A 42-year-old male presented with an unruptured middle cerebral artery bifurcation aneurysm. Device sizing was planned using high-resolution 3D DSA and volumetric analysis on a Siemens ARTIS icono biplane (ARTIS icono VE20 software; Siemens Healthineers, Erlangen, Germany). A WEB SL 5 × 3 mm device (Terumo Neuro, Tustin, CA, USA) was deployed and detached. However, the subsequent withdrawal of an adjunctive microcatheter from a branch vessel caused vascular recoil, leading the detached WEB to rotate and protrude toward the branch orifice, compromising distal flow. Instead of rescue stenting, we performed a "pushing technique" using a Leonis Mova microcatheter (SB-Kawasumi Laboratories, Kanagawa, Japan). The device was successfully reseated within the aneurysm sac, albeit with a 90° rotation. Final angiography confirmed stable seating and excellent branch patency. The patient was discharged without neurological deficits on single antiplatelet therapy. CONCLUSION: Post-detachment WEB displacement can occur due to dynamic vascular shifts even with precise volumetric sizing. Microcatheter manipulation is a feasible alternative bailout option that preserves the low-profile advantages of the WEB device and avoids the risks associated with rescue stenting.
No takes yet. Share an insight, caveat, or question.
Seo et al. (2026) conducted a case report in Unruptured left middle cerebral artery wide-neck bifurcation aneurysm with post-detachment Woven EndoBridge (WEB) device displacement (n=1). Mechanical manipulation via microcatheter and microguidewire ('pushing technique') after WEB SL 5 × 3 mm deployment was evaluated on Successful reseating of the displaced WEB device and restoration of parent branch vessel flow without rescue stenting. Microcatheter manipulation via the 'pushing technique' successfully reseated a displaced WEB device (with a 90° rotation) and restored branch flow in 1 patient, avoiding unplanned rescue stenting.
Synapse has enriched 10 related papers. Consider them for comparative context: