Detector shielding is essential in prompt gamma neutron activation analysis (PGNAA) systems to attenuate neutrons reaching the detector and suppress gamma-ray background. A Monte Carlo simulation coupled with an optimization algorithm was used to optimize the detector shielding and detector-to-sample distance. Among 35 evaluated candidates, 8 cm Pb and 5 cm BPE at 33 cm yielded the highest simulated SNR of 0.209. Experiments with the same shield also gave the highest SNR of 0.157 at 33 cm, 185.5% higher than at 25 cm. The agreement between simulation and experiment demonstrates the effectiveness of the method.