For computed tomography (CT) at fixed image noise, fixed object contrast, and a conventional reconstruction, the radiation dose required rises as the inverse fourth power of the linear voxel dimension. Halving an isotropic voxel therefore costs a factor of 16 in dose, and quartering it costs a factor of 256. With a stated reference scan of 10 mGy CTDIvol at 1.0 mm isotropic voxels, equal-noise imaging at 0.5 mm would require 160 mGy and at 0.25 mm would require 2560 mGy (2.56 Gy). Such doses are not clinically usable. This result, which follows the classical treatment of Brooks and Di Chiro, explains why improvements in clinical CT spatial resolution have come principally from detector efficiency, focal-spot and geometry improvements, and reconstruction algorithms that change the noise-resolution trade-off, rather than from increasing dose.