Answers are grounded in this paper and cite their sources.
Starting from the standard Hoult-Richards SNR relation (SNR proportional to field strength to a regime-dependent power, times voxel volume, times the square root of acquisition time over bandwidth), this paper computes the actual exchange rates among field strength, resolution, and time in MRI. In the body-noise-dominated regime that applies clinically, 1.5 T to 3 T doubles SNR and 7 T gives a factor 4.67 over 1.5 T, while one isotropic halving of voxel size divides voxel volume by 8 and multiplies the time needed at fixed SNR by 64. A 5-minute 1 mm isotropic protocol at 3 T becomes 320 minutes at 0.5 mm isotropic, and 58.8 minutes even at 7 T. The full 1.5 T to 7 T increase offsets only 0.74 of one isotropic halving. All numbers are generated by the archived computation script included with the record.
No takes yet. Share an insight, caveat, or question.
William Alexander Aldred (2026) studied this question.