RAD7 is a core instrument for geological onsite radon measurements. When measuring the exhalation rate of radon from surface soil using closed accumulation chambers, concurrently exhaled carbon dioxide accumulates within the chamber. Similarly, during in-situ degassing for water-borne radon extraction, high concentrations of carbon dioxide in water are released together with radon. Under these conditions, CO 2 concentrations typically reach tens of thousands of ppm. To evaluate this conventional field concentration range's impact on RAD7, this study focused on CO 2 interference below 50000 ppm, established a closed-loop circulation experimental system with a dual-machine parallel comparison function. RAD7-4101 acted as the reference instrument measuring CO 2 -free radon; RAD7-4103 acted as the experimental instrument, undergoing six dynamic tests across five CO 2 gradients (10000 – 50000 ppm). Paired-ratio analysis revealed an overall negative offset but no statistically detectable dependence on CO 2 concentration, consistent with the small response predicted by existing models. At 50,000 ppm, the projected incremental interference could not be statistically resolved within the experimental precision, indicating no detectable concentration-dependent CO 2 -induced suppression under the present conditions.