Answers are grounded in this paper and cite their sources.
This paper presents the design, fabrication and performance evaluation of a low-cost, locally fabricated digital RLC meter integrated with an oscilloscope interface for real-time waveform analysis.The instrument uses an Arduino Pro Mini (ATmega328P) microcontroller with its built-in 10-bit analogue-to-digital converter (ADC) to measure resistance, inductance and capacitance using voltage-divider and oscillator-frequency techniques.The system incorporates analogue signal conditioning, a three-way component-selection switch and a 2 × 16-character LCD for display of measured values, while the oscilloscope interface allows observation of voltage waveforms and phase relationships during component testing.The prototype was evaluated against seven resistors, eight capacitors and eight inductors of known nominal value (mean of three readings each) and against 17 further components of unknown value.Agreement with nominal values gave a mean absolute percentage error (MAPE) of 3.12 % for resistance (maximum 8.33%), 0.62% for capacitance (maximum 1.00 %) and 1.97 % for inductance (maximum 4.88%); all capacitors and inductors fell within their stated tolerances, and six of seven resistors fell within ± 5 %.Resistance was read systematically low by a near-constant offset (mean bias -3.12 %, 95% CI -5.49 % to -0.75%, p = 0.018), while capacitance and inductance showed no statistically significant bias (p = 0.52 and 0.70 respectively).A theoretical uncertainty budget indicates that
No takes yet. Share an insight, caveat, or question.
O. et al. (2026) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: