Answers are grounded in this paper and cite their sources.
BACKGROUND: Quantitative measurement of serum β-hydroxybutyrate (β-OHB) is essential for detecting and monitoring ketosis and ketoacidosis in diabetic cats, particularly in the era of sodium-glucose cotransporter two (SGLT2) inhibitors, where euglycemic ketoacidosis may occur. The BT3500 analyzer has not been analytically validated for β-OHB quantification in feline serum. OBJECTIVES: Analytically validate the Randox Ranbut enzymatic β-OHB assay on feline serum using the BT3500 chemistry analyzer, including evaluation of sample stability under refrigerated and frozen storage conditions. METHODS: Residual feline serum samples and pooled sera were used to assess intra-assay and inter-assay precision, spike-recovery accuracy across five concentration increments, linearity under serial dilution, and analytical sensitivity (LoB, LoD, LoQ). Short-term stability was evaluated at 4°C for 7 days and long-term stability at -20°C at predefined time points up to 360 days. LoB, LoD, and LoQ were 0.044, 0.091, and 0.150 mmol/L, respectively. β-OHB was stable at 4°C up to 6 days and at -20°C up to 180 days, while decreases exceeding the acceptance limit were observed at 360 days. CONCLUSIONS: The Randox β-OHB assay on the BT3500 demonstrated acceptable analytical performance in feline serum and is suitable for routine clinical use; frozen storage beyond 180 days may impact result accuracy.
No takes yet. Share an insight, caveat, or question.
Tirolo et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: