Key result
Dynamic bubble trap cuts arterial microbubbles by >90% during cardiopulmonary bypass in two cases.
Why the study?
Air microbubbles passing to the cerebral circulation during cardiopulmonary bypass can contribute to postoperative neurologic dysfunction, and a dynamic bubble trap was developed to diminish microbubbles in the arterial line of extracorporeal circulation.
Does a dynamic bubble trap reduce the number of large air microbubbles in the arterial line in patients undergoing coronary artery bypass grafting with cardiopulmonary bypass?
Case Report (n=2)
Does a dynamic bubble trap reduce the number of large air microbubbles in the arterial line in patients undergoing coronary artery bypass grafting with cardiopulmonary bypass?
Answers are grounded in this paper and cite their sources.
Adding a dynamic bubble trap to the arterial line during cardiopulmonary bypass markedly reduced the passage of large air microbubbles (>40 microns) despite the presence of a standard 40-micron arterial filter in two CABG patients.
Microemboli passing to the cerebral circulation during cardiopulmonary bypass can contribute to postoperative neurologic dysfunction. Many studies conclude that air microbubbles predominantly are responsible for this problem. A dynamic bubble trap (DBT) was developed to diminish the number of microbubbles in the arterial line of extracorporeal circulation. The DBT is able to substantially reduce the number of air microbubbles, as shown in two patients undergoing coronary artery bypass grafting, where a high number of microbubbles was assessed. Although a 40-micron arterial filter was used, many bubbles larger than 40 microns occurred in the arterial line. The DBT reduced the number of large microbubbles from 2,267 to 67 in patient 1 and from 897 to 61 in patient 2.
No takes yet. Share an insight, caveat, or question.
̈nburg et al. (2000) conducted a case report in Coronary artery disease undergoing coronary artery bypass grafting with cardiopulmonary bypass (n=2). Dynamic bubble trap (DBT) in the arterial line of extracorporeal circulation vs. Pre-trap arterial line with a 40-micron arterial filter was evaluated on Number of large air microbubbles (>40 microns) in the arterial line. A dynamic bubble trap reduced large microbubbles (>40 microns) during cardiopulmonary bypass from 2,267 to 67 in patient 1 and from 897 to 61 in patient 2.