21 Peak reproducibility is a major factor that affects the reliability of Fourier transform 22 ion cyclotron resonance mass spectrometry (FT-ICR MS) spectra of dissolved organic 23 matter (DOM), which was addressed by a newly proposed workflow, namely parallel 24 samples representation by similar samples (PSRSS). The performance of the workflow was 25 validated by parallel measurements and experimental replicates. Inspection of isotopic and 26 homologous peaks in the PSRSS workflow enabled elimination of electronic noise peaks 27 and low-abundance peaks with few occurrences, significantly decreased the Bray-Curtis 28 dissimilarity, and improved the molecular formula assignment ratio with only minor effects 29 on the assigned formulae. The Bray-Curtis dissimilarity threshold (55% to 60% or quartile) 30 was proposed to cluster FT-ICR MS spectra for DOM samples with similar molecular 31 composition. The noise peak removal performance was further improved for the similar 32 molecular composition samples by excluding outlier spectra and less frequently occurring 33 peaks via the PSRSS workflow. Moreover, the reliability of DOM molecular 34 characterization was significantly improved by the PSRSS workflow. Overall, our findings 35 demonstrated the feasibility of the PSRSS workflow to serve as a reliable, labor-saving 36 method to remove electronic noise peaks and cluster spectra, yielding reliable spectral 37 information for multiple independent FT-ICR MS spectra. 38
No takes yet. Share an insight, caveat, or question.