Cylindrical gear flowmeters are high-precision positive-displacement instruments widely used in aerospace, chemical engineering, and other fields requiring strict dynamic metering. However, research on their dynamic characteristics under fluctuating flow conditions remains insufficient, and conventional static calibration cannot reflect real operating performance. To address these gaps, this paper takes a DN16 cylindrical gear flowmeter as the research object and proposes a computational fluid dynamics simulation method integrated with a rotation (one degree of freedom) model. The time-varying inlet velocity is controlled by function expressions defined in the built-in C-Expression. Simulations are conducted under step flow and pulsating flow conditions, and the method is validated via real-flow experiments. Results indicate that the flowmeter behaves as a first-order inertial system under step flow. The time constant decreases with increasing step amplitude, and the response lag is more pronounced under negative step excitation. Under pulsating flow, the measured amplitude decays significantly with rising frequency; the maximum attenuation exceeds 70% within 10–50 Hz, and attenuation is more severe at low flow rates. The phase difference increases, up to 140°, with increasing frequency. This study provides a technical foundation for dynamic calibration and error compensation of cylindrical gear flowmeters.