Answers are grounded in this paper and cite their sources.
The filtering and analyzing quadrupoles of a commercial triple quadrupole mass spectrometer were modified for digital operation by replacing the sinusoidal RF drivers with high-precision digital waveform drivers. This modification enables access to higher-order Mathieu-space stability zones (i.e., zones (3,1) and (3,2)) through duty-cycle manipulation of the digital waves. To facilitate this modification, secondary digital waveforms were added to the short rods preceding the analyzing quadrupole to enhance ion transmission. These modifications provide a tractable means to expand the operating parameters and potentially the m/z range of commercial triple quadrupole systems.
No takes yet. Share an insight, caveat, or question.
Groetsema et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: