Key result
The Pulsatile Ventricular Assist Platform in VA-ECMO (n=6 pigs) increased forward cardiac output without LV overload, and synchronized counter-pulsatile flow improved LV afterload vs no support.
Why the study?
Continuous-flow VA-ECMO temporarily maintains hemodynamics at the cost of overloading the left ventricle, whereas a new Pulsatile Ventricular Assist Platform could offer more physiological hemodynamic support.
Does a Pulsatile Ventricular Assist Platform (pVAP) with synchronized counter-pulsatile flow in a VA-ECMO configuration improve systemic and intraventricular hemodynamics compared to continuous-flow support and no support in porcine models of acute ischemic left ventricular failure?
Does a Pulsatile Ventricular Assist Platform (pVAP) with synchronized counter-pulsatile flow in a VA-ECMO configuration improve systemic and intraventricular hemodynamics compared to continuous-flow support and no support in porcine models of acute ischemic left ventricular failure?
Answers are grounded in this paper and cite their sources.
In a porcine model of acute ischemic heart failure, a novel Pulsatile Ventricular Assist Platform in a VA-ECMO configuration provided stable hemodynamic support and improved left ventricular afterload and ventricular-arterial coupling when using synchronized counter-pulsatile flow.
Background and aims Mortality from heart failure is still high despite all the recent advances. Continuous-flow (CF) Veno-arterial Extracorporeal Membrane Oxygenation (VA-ECMO) can temporarily maintain systemic and pulmonary hemodynamics, at the cost of overloading the left ventricle (LV). A new Pulsatile Ventricular Assist Platform (pVAP) could offer a more physiological hemodynamic support. Methods 6 large porcine models of acute ischemic left ventricular (LV) failure received the pVAP in a VA-ECMO configuration. Invasive systemic and intraventricular hemodynamic measurements were taken with synchronized counter-pulsatile flow (SCPF) and CF support at the same flow rate. The measurements were compared with no support (ON vs OFF) in both modalities. Results The pVAP successfully delivered VA-ECMO in acute ischemic heart failure. The device maintained stable hemodynamics and increased the forward cardiac output (CO F ) without overloading the LV and with no change in the LV pressure-volume area (PVA). SCPF, but not CF, was associated with improved LV afterload and ventricular-arterial coupling compared to no support. Conclusion The findings suggest that the pVAP is feasible also in a VA-ECMO configuration, supporting hemodynamics with more physiological effect on effect on the LV dynamics.
No takes yet. Share an insight, caveat, or question.
Bastos et al. (2026) studied Acute ischemic left ventricular failure (n=6). Pulsatile Ventricular Assist Platform (pVAP) with synchronized counter-pulsatile flow (SCPF) in VA-ECMO configuration vs. Continuous-flow (CF) support and no support (OFF) was evaluated on Invasive systemic and intraventricular hemodynamic measurements (forward cardiac output, LV pressure-volume area, LV afterload, and ventricular-arterial coupling). The Pulsatile Ventricular Assist Platform in VA-ECMO (n=6 pigs) increased forward cardiac output without LV overload, and synchronized counter-pulsatile flow improved LV afterload vs no support.