We present the design, fabrication, and characterization of an aluminum-based six-channel microwave SQUID multiplexer ( μ MUX) prototype for transition-edge sensor readout in the Ricochet experiment. The device consists of aluminum coplanar-waveguide resonators and RF superconducting quantum interference devices with Dolan-style Al/AlO x /Al Josephson junctions. By measuring the resonator scattering parameters at a range of probe tone frequencies, powers, and flux bias points, we demonstrate agreement between the device response and existing multiplexer models. We also characterize the noise performance in both open-loop and flux-ramping modes. With a high electron mobility transistor (HEMT) amplifier, open-loop measurements yield a flux sensitivity of 1–1.5 μ Φ 0 / Hz . With flux-ramp modulation, low-frequency 1 / f noise is suppressed, and the flux sensitivity is around 3–4 μ Φ 0 / Hz , corresponding to a current sensitivity of 24–33 pA / Hz at the input coil. We further demonstrate a reduction in readout noise by incorporating a Josephson traveling-wave parametric amplifier between the μ MUX and the HEMT. This achieves an open-loop flux sensitivity of 0.3–0.6 μ Φ 0 / Hz and an effective system noise temperature below 1K. These results establish aluminum μ MUX devices as a viable and extensible readout technology for low-noise cryogenic detector arrays.